Measurement of retinal thickness from three-dimensional images obtained from C scan images from the optical coherence tomography ophthalmoscope.

机译:根据从光学相干断层扫描检眼镜的C扫描图像获得的三维图像测量视网膜厚度。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: One of the unique features of the optical coherence tomography (OCT) ophthalmoscope is that it can record C scan images of the retina. The purpose of this study was to determine the best recording time to measure the retinal thickness with the OCT ophthalmoscope. In addition, the accuracy of the measurements was examined by comparing the values obtained by the OCT ophthalmoscope with those obtained with the Stratus OCT Model 3000 (OCT III) assuming that the OCT III gives an accurate measurement of retinal thickness. METHODS: The topography mode of the OCT ophthalmoscope was used. First, the average retinal thickness recorded with 2-s scans was compared with that recorded with 4-s scans for a recording area of 15 degreesx15 degrees. Next, the average retinal thickness recorded by the OCT ophthalmoscope was compared with that obtained by the Fast Macular Thickness program of the OCT III in patients with macular oedema. RESULTS: The mean retinal thickness of the central area was 208.1 microm for both 2 and 4 s recording times. The average retinal thickness obtained by the OCT ophthalmoscope was highly correlated and not significantly different from the values obtained by the OCT III. The largest differences obtained by the two instruments were seen in the parts of the retina with accumulation of hard exudates. CONCLUSIONS: Accurate measurements of retinal thickness can be obtained with the OCT ophthalmoscope by 2-s scans, and thus, the OCT ophthalmoscope can be a valuable instrument for clinical assessments of retinal thickness.
机译:背景技术:光学相干断层扫描(OCT)检眼镜的独特特征之一是它可以记录视网膜的C扫描图像。这项研究的目的是确定使用OCT检眼镜测量视网膜厚度的最佳记录时间。此外,假设OCT III可以准确测量视网膜厚度,则可以通过比较OCT检眼镜的测量值与Stratus OCT Model 3000(OCT III)的测量值来检查测量的准确性。方法:使用OCT检眼镜的地形模式。首先,对于15度x 15度的记录区域,将2秒扫描记录的平均视网膜厚度与4秒扫描记录的平均视网膜厚度进行比较。接下来,将OCT检眼镜记录的平均视网膜厚度与通过OCT III的快速黄斑厚度程序获得的黄斑水肿患者的平均视网膜厚度进行比较。结果:在记录的2和4 s时间内,中心区域的平均视网膜厚度为208.1微米。 OCT检眼镜检出的平均视网膜厚度与OCT III检出的高度相关且无显着差异。两种仪器获得的最大差异出现在视网膜部分,伴有硬性渗出物积聚。结论:OCT检眼镜可以通过2-s扫描获得准确的视网膜厚度测量值,因此,OCT检眼镜可以成为临床评估视网膜厚度的有价值的工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号