Reliability of low platelet counts--comparison of microscopic counts with automated hematology analyzer XE-2100

机译:低血小板计数的可靠性-与自动血液分析仪XE-2100进行的显微计数比较

获取原文
获取原文并翻译 | 示例
           

摘要

Sysmex XE-2100 is an automated hematological analyzer with excellent features that can count platelets by both impedance and optical fluorescent method. Particularly, it is notable that platelet counts by optical fluorescent method (PLT-O) are automatically adopted when platelet counts by impedance method (PLT-I) are less than 50 x 10(3)/microl or platelets have abnormal distribution. We compared PLT-I with PLT-O, and also with microscopic counts, using the samples with thrombocytopenia. Regarding reproducibility, mean coefficient of variation values were 5.3% in both PLT-I and PLT-O, when no flags of "PLT Abn Distribution" appeared in the samples with thrombocytopenia. Coefficient of variation values was 10.4% in PLT-I and 5.9% in PLT-O, when flags of "PLT Abn Distribution" appeared in the sample with thrombocytopenia. Correlation among the data obtained by PLT-I, PLT-O and microscopic counts were excellent. When a large difference was observed between PLT-I and PLT-O, PLT-O was more closely consistent with microscopic counts. PLT-O was useful compared with PLT-I, and this system was considered to measure platelet counts more correctly in the samples with thrombocytopenia. Thus, it was suggested that reliable platelet counts could be reported even in the samples with thrombocytopenia by switching system in XE-2100.
机译:Sysmex XE-2100是一款自动血液分析仪,具有出色的功能,可以通过阻抗法和光学荧光法计数血小板。特别地,值得注意的是,当通过阻抗法(PLT-1)的血小板计数小于50×10(3)/μl或血小板具有异常分布时,自动采用通过光学荧光法(PLT-O)的血小板计数。我们使用血小板减少症样本将PLT-1与PLT-0以及显微镜计数进行了比较。关于重现性,当血小板减少症样本中没有出现“ PLT Abn分布”的标志时,PLT-1和PLT-0的平均变异系数值为5.3%。当血小板减少症样本中出现“ PLT Abn分布”标志时,PLT-1中变异系数的值为10.4%,PLT-O中变异系数的值为5.9%。通过PLT-1,PLT-0获得的数据与显微计数之间的相关性极好。当在PLT-1和PLT-O之间观察到很大的差异时,PLT-O与显微计数更加一致。与PLT-1相比,PLT-0是有用的,并且该系统被认为可以更正确地测量血小板减少症样本中的血小板计数。因此,有人建议通过XE-2100的切换系统,即使在血小板减少症的样本中也可以报告可靠的血小板计数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号