首页> 外文期刊>Journal of medical imaging and radiation oncology >Correlation between CT attenuation value and iodine concentration in vitro: Discrepancy between gemstone spectral imaging on single-source dual-energy CT and traditional polychromatic X-ray imaging
【24h】

Correlation between CT attenuation value and iodine concentration in vitro: Discrepancy between gemstone spectral imaging on single-source dual-energy CT and traditional polychromatic X-ray imaging

机译:体外CT衰减值与碘浓度的相关性:单源双能CT上的宝石光谱成像与传统的多色X射线成像之间的差异

获取原文
获取原文并翻译 | 示例
           

摘要

Introduction: To assess the relation between CT attenuation value and iodine concentration in vitro, using gemstone spectral imaging (GSI) with single-source dual-energy CT and traditional polychromatic X-ray imaging (TPXI), respectively. Methods: A polypropylene phantom with eight test tubes in which iodine concentrations of solution were 0.4, 0.7, 2, 5, 10, 20, 30 and 50 mg/mL underwent GSI and traditional polychromatic X-ray scans (80, 100, 120 and 140 kV(p)), using single-source dual-energy spectral CT (Discovery CT750HD; GE Healthcare Technologies, Milwaukee, WI, USA) at the same tube speed of 0.8 s/rotation. All spectral imaging data were analysed with GSI viewer to obtain monochromatic images (50-140 keV, interval of 10 keV). Computed tomography attenuation values of iodine solution were measured with the same size of regions of interest and at the exact same level for both monochromatic and polychromatic images. The relation between CT attenuation value and iodine concentration was examined. Results: A linear correlation was found between CT attenuation value and iodine concentration for both monochromatic and polychromatic images. Moreover, the fitting coefficients for CT attenuation values and iodine concentrations were closer to one with GSI (r2 = 0.99824-0.99996) than that with TPXI (r2 = 0.99640-0.99736). Conclusions: Owing to the better correlation coefficients between CT attenuation value and iodine concentration, GSI may be a preferred method for quantitative measurement compared with TPXI.
机译:简介:为了评估体外CT衰减值与碘浓度之间的关系,分别使用具有单源双能CT的宝石光谱成像(GSI)和传统的多色X射线成像(TPXI)。方法:采用八个试管的聚丙烯模型,其中碘溶液浓度分别为0.4、0.7、2、5、10、20、30和50 mg / mL进行GSI和传统的多色X射线扫描(80、100、120和使用单源双能谱CT(Discovery CT750HD; GE Healthcare Technologies,美国威斯康星州密尔沃基),在相同的管速为0.8 s /旋转的情况下达到140 kV(p))。使用GSI Viewer分析所有光谱成像数据以获得单色图像(50-140 keV,间隔10 keV)。在单色区域和多色图像的感兴趣区域的大小相同且水平完全相同的情况下,测量了碘溶液的计算机断层扫描衰减值。检查了CT衰减值与碘浓度之间的关系。结果:对于单色和多色图像,CT衰减值与碘浓度之间存在线性关系。而且,GSI(r2 = 0.99824-0.99996)的CT衰减值和碘浓度的拟合系数比TPXI(r2 = 0.99640-0.99736)的系数更接近1。结论:由于CT衰减值与碘浓度之间具有更好的相关系数,所以与TPXI相比,GSI可能是一种首选的定量测量方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号