Significant time until catheter occlusion alerts in currently marketed insulin pumps at two basal rates

机译:当前市售胰岛素泵中以两种基本速率出现导管阻塞之前需要花费大量时间

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

One of the most common problems of continuous subcutaneous insulin infusion (CSII) is occlusion of the infusion set. Frequencies up to 30% have been reported in literature, although objective assessment of occlusion is difficult. CSII devices feature an occlusion detection alert based on a sensor connected to an electronics system that measures pressure, compression, or displacement of the lead screw in the pump drive system. Above a certain threshold, an alert is given. A relatively large displacement of the lead screw may be needed to activate the alert; in other words, it may take quite some time before the occlusion alert is activated. Some CSII manufacturers provide information in the manual on the time needed to activate the alert after total occlusion. For example, the time until occlusion alert in the D-Tron~(TM) plus (Disetronic Medical Systems, Member of the Roche Group, Basel, Switzerland) is given as between 2 and 4h at a basal rate of 1.0IU/h. If the basal rate decreases, the occlusion alert will be activated later. As an illustration, the time until occlusion alert of the Accu-Chek~R Spirit Combo (Disetronic Medical Systems) is given as shorter than 5 h at a basal rate of 1.0IU/h, whereas at a basal rate of 0.1 IU/h, the alert will go off after up to 50 h. An occlusion alert is especially important during the night; therefore we examined the time until occlusion alert at basal rates.
机译:连续皮下胰岛素输注(CSII)的最常见问题之一是输注装置的堵塞。尽管很难客观评估阻塞的发生率,但文献报道的频率高达30%。 CSII设备具有闭塞检测警报功能,该警报基于连接到电子系统的传感器,该传感器测量泵驱动系统中导螺杆的压力,压缩或位移。超过某个阈值,将发出警报。可能需要导螺杆相对较大的位移才能激活警报。换句话说,可能需要花费一些时间才能激活遮挡警报。一些CSII制造商在手册中提供了有关在完全堵塞后激活警报所需时间的信息。例如,以1.0IU / h的基准速率给出直到D-Tron TM Plus(Disetronic Medical Systems,罗氏集团成员,瑞士巴塞尔)中的闭塞警报的时间为2-4小时。如果基础速率降低,则稍后会激活阻塞警报。作为说明,以Accu-Chek〜R Spirit Combo(Disetronic Medical Systems)的闭塞警报之前的时间以1.0IU / h的基础速率小于5小时,而以0.1 IU / h的基础速率,最多50小时后警报将响起。夜间阻塞警报尤为重要;因此,我们以基础速率检查了直到发生闭塞警报的时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号