首页> 外文期刊>Journal of thermal analysis and calorimetry >Absolute measurement of thermal conductivity of poly (acrylic acid) by transient hot wire technique
【24h】

Absolute measurement of thermal conductivity of poly (acrylic acid) by transient hot wire technique

机译:瞬态热丝技术绝对测量聚丙烯酸的导热系数

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The transient hot-wire method is considered the most accurate technique to measure the thermal conductivity of fluids. In this study, a transient hot wire instrument which employs 25.4-μm-diameter tantalum wire with an insulating tantalum pentoxide coating has been used. This hot-wire cell with a thin insulating layer is suitable for measurement of the thermal conductivity of electrically conducting and polar liquids. Measurements of the thermal conductivity of 50 wt% solution of PAA [poly (acrylic acid)] in water and PAA-Na in 50 wt% water are reported here. These measurements were obtained in the temperature range of 299-368 K at 1 atmospheric pressure. The measurement of thermal conductivity is estimated to be accurate within ±4%.
机译:瞬态热线法被认为是测量流体热导率的最准确技术。在这项研究中,已经使用了瞬态热线仪器,该仪器采用直径为25.4μm的钽丝并带有五氧化二钽绝缘涂层。这种具有薄绝缘层的热线电池适用于测量导电液体和极性液体的热导率。此处报告了50%(重量)的PAA [聚(丙烯酸)]水溶液和PAA-Na在50%(重量)的水溶液中的导热率的测量结果。这些测量是在1个大气压下于299-368 K的温度范围内获得的。热导率的测量估计准确度在±4%以内。