首页> 外文期刊>Journal of the mechanical behavior of biomedical materials >Embrittlement phenomenon of Ag core MP35N cable as lead conductor in medical device
【24h】

Embrittlement phenomenon of Ag core MP35N cable as lead conductor in medical device

机译:Ag芯MP35N电缆作为医疗器械引线导体的脆化现象

获取原文
获取原文并翻译 | 示例
           

摘要

Ag core MP35N (Ag/MP35N) wire has been used in lead electric conductor wires in the medical device industry for many years. Recently it was noticed that the combination of silver and MP35N restricts its wire drawing process. The annealing temperature in Ag/MP35N has to be lower than the melting temperature of pure Ag (960. ??C), which cannot fully anneal MP35N. The lower annealing temperature results in a highly cold worked MP35N, which significantly reduces Ag/MP35N ductility. The embrittlement phenomenon of Ag/MP35N cable was observed in tension and bending deformation. The effect of the embrittlement on the wire flex fatigue life was evaluated using a newly developed flex fatigue testing method. The Ag/MP35N cable fatigue results was analyzed with a Coffin-Manson approach and compared to the MP35N cable fatigue results. The root causes of the Ag/Mp35N embrittlement phenomenon are discussed. ? 2012 Elsevier Ltd.
机译:银芯MP35N(Ag / MP35N)导线已在医疗设备行业中用作导线导线。最近,人们注意到银和MP35N的结合会限制其拉丝工艺。 Ag / MP35N中的退火温度必须低于不能完全退火MP35N的纯Ag的熔化温度(960°C)。较低的退火温度会导致MP35N经过高度冷加工,从而大大降低了Ag / MP35N的延展性。在拉伸和弯曲变形中观察到Ag / MP35N电缆的脆化现象。使用新开发的挠曲疲劳测试方法评估了脆化对金属丝挠曲疲劳寿命的影响。使用Coffin-Manson方法分析了Ag / MP35N电缆的疲劳结果,并将其与MP35N电缆的疲劳结果进行了比较。讨论了Ag / Mp35N脆化现象的根本原因。 ? 2012爱思唯尔有限公司