首页> 外文期刊>Journal of the mechanical behavior of biomedical materials >Late failure of annealed highly cross-linked polyethylene acetabular liner
【24h】

Late failure of annealed highly cross-linked polyethylene acetabular liner

机译:退火的高度交联聚乙烯髋臼内衬的晚期失效

获取原文
获取原文并翻译 | 示例
           

摘要

Highly cross-linked polyethylene (HXPE) in total hip arthroplasty (THA) has been shown to significantly decrease wear rates compared with conventional polyethylene (CPE). However, crosslinking, thermal treatment and oxidation can decrease the mechanical properties of PE, and several cases of fracture of remelted HXPE liners were reported. We present, for the first time, unexpected failures of THA with the use of annealed HXPE liners in two patients occurring at 7 and 8 years after operation. Operative findings revealed dislocated liners from the metal shell and a fracture of the superior rim at the rim-dome junction in both liners. Scanning electron microscopy showed that the cracks initiated at the rim and propagated toward the articular surface. Both liners showed generally a low amount of oxidation (less than 1.00) at the articular surface and low wear rates; however, oxidation at the rim was relatively higher (mean 1.55). These findings suggested that decreased mechanical properties at the rim-dome junction due to cross-linking, annealing and oxidation might have been caused breakage of the HXPE liners after a long implantation time, although the annealed HXPE achieved low degree of wear.
机译:与常规聚乙烯(CPE)相比,全髋关节置换术(THA)中高度交联的聚乙烯(HXPE)已显示出可显着降低磨损率。但是,交联,热处理和氧化会降低PE的机械性能,并且有几起重熔的HXPE内衬断裂的报道。我们首次对两名患者在手术后7年和8年使用退火的HXPE衬管呈现出THA的意外失败。手术结果显示,衬板从金属壳中脱位,并且两个衬板的轮辋-圆顶接合处上轮辋均断裂。扫描电子显微镜显示,裂纹从边缘开始并向关节表面传播。两种衬里在关节表面通常都表现出少量的氧化(小于1.00)和低磨损率。但是,轮辋的氧化程度相对较高(平均值为1.55)。这些发现表明,尽管退火后的HXPE磨损程度较低,但是由于交联,退火和氧化作用导致的边缘-穹顶接合处机械性能下降可能是由于长时间的植入后HXPE衬里破裂而引起的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号