Comparison of the Performance between Portal Dosimetry and a Commercial Two-dimensional Array System on Pretreatment Quality Assurance for Volumetric-modulated Arc and Intensity-modulated Radiation Therapy

机译:门控剂量学和商用二维阵列系统在容积调制电弧和强度调制放射治疗的预处理质量保证上的性能比较

获取原文
获取原文并翻译 | 示例
       

摘要

The aim of this study was to compare the dosimetric performance and to evaluate the pretreatment quality assurance (QA) of a portal dosimetry and a commercial two-dimensional (2-D) array system. In the characteristics comparison study, the measured values for the dose linearity, dose rate response, reproducibility, and field size dependence for 6-MV photon beams were analyzed for both detector systems. To perform the qualitative evaluations of the 10 IMRT and the 10 VMAT plans, we used the Gamma index for quantifying the agreement between calculations and measurements. The performance estimates for both systems show that overall, minimal differences in the dosimetric characteristics exist between the Electron portal imaging device (EPID) and 2-D array system. In the qualitative analysis for pretreatment quality assurance, the EPID and 2-D array system yield similar passing rate results for the majority of clinical Intensity-modulated radiation therapy (IMRT) and Volumetric-modulated arc therapy (VMAT) cases. These results were satisfactory for IMRT and VMAT fields and were within the acceptable criteria of γ%≤1, γ_(avg)<0.5. The EPDI and the 2-D array systems showed comparable dosimetric results. In this study, the results revealed both systems to be suitable for patient-specific QA measurements for IMRT and VMAT. We conclude that, depending on the status of clinic, both systems can be used interchangeably for routine pretreatment QA.
机译:这项研究的目的是比较剂量学性能并评估门诊剂量学和商业二维(2-D)阵列系统的预处理质量保证(QA)。在特性比较研究中,对两个探测器系统的6-MV光子束的剂量线性,剂量率响应,重现性和视场大小依赖性的测量值进行了分析。为了对10个IMRT和10个VMAT计划进行定性评估,我们使用Gamma指数来量化计算和测量之间的一致性。两种系统的性能评估表明,电子门成像设备(EPID)与二维阵列系统之间的剂量特性总体上存在最小差异。在进行预处理质量保证的定性分析中,对于大多数临床强度调制放射治疗(IMRT)和容积调制弧光治疗(VMAT)病例,EPID和2-D阵列系统得出的合格率结果相似。这些结果对于IMRT和VMAT领域是令人满意的,并且在γ%≤1,γ_(avg)<0.5的可接受标准内。 EPDI和二维阵列系统显示出可比的剂量学结果。在这项研究中,结果表明这两种系统均适用于IMRT和VMAT的患者特定QA测量。我们得出的结论是,根据临床状况,两种系统均可互换用于常规预处理QA。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号