首页> 外文期刊>Journal of the Korean Physical Society >Dual-source Parallel Radiofrequency Excitation ACR Phantom Magnetic Resonance Imaging at 3 T: Assessment of the Effect of Image Quality on High-contrast Spatial Resolution, Percent Signal Ghosting, and Low-contrast Object Detectability in Comparison with Conventional Single-source Transmission
【24h】

Dual-source Parallel Radiofrequency Excitation ACR Phantom Magnetic Resonance Imaging at 3 T: Assessment of the Effect of Image Quality on High-contrast Spatial Resolution, Percent Signal Ghosting, and Low-contrast Object Detectability in Comparison with Conventional Single-source Transmission

机译:3 T时双源并行射频激励ACR幻影磁共振成像:与传统单源传输相比,评估图像质量对高对比度空间分辨率,信号重影百分比和低对比度对象可检测性的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of the present study was to assess dual-source parallel radiofrequency (RF) excitation American College of Radiology (ACR) phantom magnetic resonance (MR) imaging at 3T compared with conventional single-source RF transmission and compared with the standard ACR MRI phantom test. We used a 3T MR scanner equipped with dual-source parallel RF excitation and an 8-channel head phased array coil. We employed T1- and T2-weighted fast spin echo (FSE) pulse sequences for an assessment of the impact of image quality on high-contrast spatial resolution, percent signal ghosting and low-contrast object detectability following the ACR MRI quality control (QC) manual. With geometric accuracy and identical slice locations, dual RFs using dual-source parallel RF excitation MR showed an advantage over single RF using dual-source parallel RF excitation MR and conventional MR in terms of high-contrast spatial resolution (p < 0.010), percent signal ghosting (p < 0.010), and low-contrast object detectability (p < 0.010). The quality of the image from the dual-source parallel RF excitation MR equipment was superior to that of the image from conventional MR equipment for the ACR phantom. We need to pursue dual-source parallel RF excitation MR studies involving various clinical cases.
机译:本研究的目的是评估美国传统放射学院(ACR)幻影磁共振(MR)在3T时的双源并行射频(RF)激发与常规单源RF传输相比,并与标准ACR MRI幻像进行比较测试。我们使用的3T MR扫描仪配备了双源并行RF激励和8通道头部相控阵线圈。我们采用T1和T2加权快速自旋回波(FSE)脉冲序列来评估图像质量对ACR MRI质量控制(QC)后高对比度空间分辨率,信号重影百分比和低对比度对象可检测性的影响手册。凭借几何精度和相同的切片位置,使用双源并行RF激励MR的双RF在使用高对比度空间分辨率(p <0.010)方面显示出优于使用双源并行RF激励MR和常规MR的单RF。信号重影(p <0.010)和低对比度物体可检测性(p <0.010)。来自双源并行RF激励MR设备的图像质量优于用于ACR体模的传统MR设备的图像质量。我们需要进行涉及各种临床病例的双源并行RF激发MR研究。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号