首页> 外文期刊>Journal of the American Association for Laboratory Animal Science >Comparison of Traditional and Integrated Digital Anesthetic Vaporizers
【24h】

Comparison of Traditional and Integrated Digital Anesthetic Vaporizers

机译:传统麻醉机与数字麻醉机的比较

获取原文
获取原文并翻译 | 示例
       

摘要

Recent efforts have focused on mitigating anesthetic gas emissions during laboratory animal experiments. A recently developed, digitally controlled, integrated digital vaporizer (IDV) using a syringe pump has been designed to use and administer anesthetic gas to mice and rats more efficiently. The entire IDV system can be placed on a laboratory bench, requires fewer charcoal filters to act as passive scavengers when used at a low gas flow rate, and does not need compressed gas to operate, a requirement for traditional passive systems. The objective of this study was to compare isoflurane usage between a traditional vaporizer (TdV) and an IDV system at both the same settings and those recommended by the manufacturer. We used 10 C57BL/6 male mice and administered isofturane through either nose cones or tracheal tubes connected to a pulsatile ventilator. The results showed that isoflurane usage is highly dependent on the flow rate of the carrier gas, but the IDV system was more precise and handled low flow rates (150 mL/min) better than did the TdV system. We observed only slight differences in heart rate, respiration rate, core body temperature, time to loss of the righting reflex, and recovery time between group averages for both systems when set to manufacturer-recommended settings. Although observed decreased levels of waste anesthetic gas at low flow rates are expected from the IDV system, further work is needed to assess environmental anesthetic gas levels and exposure to laboratory personnel.
机译:最近的努力集中在减轻实验室动物实验过程中的麻醉气体排放。已经设计出使用注射泵的最近开发的数字控制的集成数字汽化器(IDV),以更有效地向小鼠和大鼠使用和施用麻醉气体。整个IDV系统可放置在实验室工作台上,以较低的气体流量使用时,需要较少的活性炭过滤器作为被动清除剂,并且不需要压缩气体即可操作,这是传统被动系统的要求。这项研究的目的是比较传统蒸发器(TdV)和IDV系统在相同设置和制造商推荐的设置下的异氟醚用量。我们使用了10只C57BL / 6雄性小鼠,并通过与脉动呼吸机相连的鼻锥或气管导管给予异氟烷。结果表明,异氟醚的使用高度依赖于载气的流速,但IDV系统比TdV系统更精确,处理低流速(150 mL / min)更好。当设置为制造商建议的设置时,两个系统的组平均值之间的心率,呼吸频率,核心体温,失去正向反射的时间以及恢复时间之间的差异仅很小。尽管可以预期IDV系统在低流速下观察到的麻醉麻醉气体水平降低,但仍需要进一步的工作来评估环境麻醉气体水平以及对实验室人员的暴露程度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号