A method for determining the earthy and musty odors 2-methylisoborneol (2-MIB) and geosmin in drinking water using static headspace-GC-MS is described. To achieve lower detection limits, split ratio was optimized with ramped inlet pressure for large headspace sampling volume. The ramped inlet pressure, which held higher pressure (higher column flow rate) only during injection, allowed us to inject 3-mL volume to GC with very low split ratio (2:1). Although sequential analysis with a stainless steel ion source often changed the mass spectrum of 2-MIB, this spectral change was eliminated by using an inert ion source with a 6 mm drawout plate. The detection limits of this method were 0.36 and 0.14 ng/L, respectively, for 2-MIB and geosmin. The repeatabilities (n = 30) were 6.6 and 4.8%, respectively, at 1 ng/L for 2-MIB and geosmin.
机译:描述了一种使用静态顶空-GC-MS测定饮用水中的土和霉味2-甲基异冰片醇(2-MIB)和土臭味的方法。为了达到较低的检测限,分流比通过增大入口压力进行了优化,以实现较大的顶空进样量。上升的入口压力仅在进样过程中才保持较高的压力(较高的色谱柱流速),这使我们能够以极低的分流比(2:1)向GC进样3-mL体积。尽管使用不锈钢离子源进行的顺序分析通常会改变2-MIB的质谱,但是通过使用带有6 mm拉出板的惰性离子源可以消除这种光谱变化。对于2-MIB和土臭素,该方法的检出限分别为0.36和0.14 ng / L。 2-MIB和土臭素在1 ng / L时的重复性(n = 30)分别为6.6和4.8%。