PURPOSE: To evaluate the effects of automating the registration process for wavefront measurement using the LADARWave aberrometer. Measures of interest include accuracy, repeatability, and speed of measurement. METHODS: Five separate registration images were collected for each of eight eyes of four individuals with varying iris color using the LADARWave aberrometer. Technicians manually aligned the images and the autoregistration feature was used to align the same images. Time, accuracy, and repeatability of the alignment were measured using objective techniques. The potential effect on positional error was evaluated using a retrospective reprocessing of LADARWave data from an existing sample dataset. RESULTS: Autoregistration was the fastest method of alignment, from five to seven times faster than manual registration. It was also the most consistent, with the lowest pupil/limbus offset error (45 microm). Technicians averaged 48 to 91 microm. The differences in wavefronts used to generate the composite wavefront image for surgery export were 18% lower with autoregistration. CONCLUSIONS: The autoregistration feature of the LADARWave aberrometer not only removes inter-technician variability, but reduces the time to perform image alignments and improves the consistency of the alignments. This helps reduce errors related to wavefront position, potentially improving outcomes.
机译:目的:评估使用LADARWave像差仪自动进行波前测量配准过程的效果。感兴趣的度量包括准确性,可重复性和测量速度。方法:使用LADARWave像差计,为四只虹膜颜色不同的个人的八只眼睛的每只眼睛收集五张单独的配准图像。技术人员手动对齐图像,并使用自动配准功能对齐相同的图像。使用客观技术测量比对的时间,准确性和可重复性。使用来自现有样本数据集的LADARWave数据的回顾性重新处理,评估了对位置误差的潜在影响。结果:自动配准是最快的对齐方式,比手动配准快五到七倍。它也是最一致的,具有最小的瞳孔/边缘偏移误差(45微米)。技术人员平均48到91微米。自动注册后,用于生成手术出口的复合波前图像的波前差异降低了18%。结论:LADARWave像差仪的自动配准功能不仅消除了技术人员之间的差异,而且减少了执行图像对准的时间,并改善了对准的一致性。这有助于减少与波前位置有关的误差,从而可能改善结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号