Estimation of uncertainty in kinetic-colorimetric assay of bacterial endotoxins.

机译:细菌内毒素动力学比色法测定的不确定性估计。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The relative standard deviation (R.S.D.) of measurements is estimated in the kinetic-colorimetric assay of bacterial endotoxins without recourse to the usual repeated experiments. The measurements are the slopes of kinetic curves and two major factors are considered to cause the uncertainty of the measurements: (1) the pipetting of the sample and color development reagent; and (2) noise in the detection unit. The measurement R.S.D. is formulated as a function of endotoxin concentration. Two parameters (S.D. of the pipetted volumes and S.D. of the detector noise) are also required in the uncertainty equation, but no arbitrary coefficients are included. Since the S.D. values for pipettes and detector noise can be determined independently of the endotoxin assays, the measurement R.S.D. can be estimated by the above equation without repeating the assays. However, the calibration curve is necessary. The theoretical estimation is shown to be in good agreement with the experimental R.S.D. (n = 12) over a wideconcentration range.
机译:测量的相对标准偏差(R.S.D.)是在细菌内毒素的动力学比色测定中估算的,无需借助通常的重复实验。测量值是动力学曲线的斜率,并且考虑到两个主要因素导致测量值的不确定性:(1)移液样品和显色试剂; (2)检测单元中的噪声。测量R.S.D.根据内毒素浓度来配制。在不确定性方程式中还需要两个参数(移液体积的S.D.和检测器噪声的S.D.),但是不包括任意系数。自S.D.移液管和检测器噪声的值可以独立于内毒素测定法(R.S.D.可以通过以上等式估算,而无需重复测定。但是,校准曲线是必需的。理论估计值与实验R.S.D. (n = 12)在宽浓度范围内。