Clinical study on the insertion torque and wireless resonance frequency analysis in the assessment of torque capacity and stability of self-tapping dental implants.

机译:插入扭矩和无线共振频率分析在评估自攻牙种植体扭矩容量和稳定性方面的临床研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Resonance frequency (RF) analysis is a non-invasive, objective and sensitive technique developed for implantology where it measures the stability of the implant in osteotomy site. Although many studies were performed by the previous electronic version of RF analyzer, a very limited number of studies were carried out with the new magnetic wireless version. The aim of the study was to evaluate the relation between insertion torques, primary and secondary stability of self-tapping tapered implant systems. Thirteen subjects were treated with 42 endosseous implants using two-stage surgical procedure. The maximal insertion torque values were recorded prior to RF analysis during surgery. Six months after surgery, the secondary stability values were measured by the RF analysis. The average maximal insertion torque and primary and secondary magnetic RF values were 33 +/- 11 N cm and 66 +/- 12 ISQ and 71.9 +/- 6 ISQ for 42 implants respectively. The correlation between insertion torque and RF values were indicated to be statistically significant (P < 0.01). Significantly higher maximal insertion torque, and primary and secondary magnetic RF values were achieved in mandibular sites compared with maxillary areas (P < 0.01). No significant differences were measured for all parameters when both systems were compared with each other (P > 0.05). There was a strong correlation between the insertion torque, primary and secondary magnetic RF values of self-tapping tapered endosseous implant used. Further studies are needed to understand the impact of the wireless magnetic RF analysis technique in clinics.
机译:共振频率(RF)分析是一种非侵入性,客观且灵敏的技术,专为种植术开发,用于测量截骨部位植入物的稳定性。尽管以前的射频分析仪的电子版本进行了许多研究,但是使用新型磁性无线版本进行的研究却非常有限。该研究的目的是评估自攻锥形植入物系统的插入扭矩,初级和次级稳定性之间的关系。使用两阶段外科手术程序,用42枚骨内植入物治疗了13名受试者。在手术期间进行RF分析之前,记录最大插入扭矩值。手术六个月后,通过RF分析测量二级稳定性值。 42个植入物的平均最大插入扭矩以及一次和二次磁性RF值分别为33 +/- 11 N cm和66 +/- 12 ISQ和71.9 +/- 6 ISQ。插入扭矩和RF值之间的相关性表明具有统计学意义(P <0.01)。与上颌骨区域相比,在下颌部位获得的最大插入扭矩明显更高,并且初级和次级磁性RF值均达到了(P <0.01)。当两个系统相互比较时,所有参数均未检测到显着差异(P> 0.05)。插入扭矩,所用自攻锥形骨质植入物的初级和次级磁性RF值之间有很强的相关性。需要进一步的研究以了解无线电磁射频分析技术在临床中的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号