首页> 外文期刊>Journal of orthopaedic research >Wear and migration of highly cross-linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: a randomized radiostereometric study of 150 patients.
【24h】

Wear and migration of highly cross-linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: a randomized radiostereometric study of 150 patients.

机译:高度交联的传统硬质聚乙烯杯和钴铬合金或Oxinium股骨头的磨损和迁移:一项150名患者的随机放射线立体测量研究。

获取原文
获取原文并翻译 | 示例

摘要

This randomized study was performed to compare wear and migration of five different cemented total hip joint articulations in 150 patients. The patients received either a Charnley femoral stem with a 22.2 mm head or a Spectron EF femoral stem with a 28 mm head. The Charnley articulated with a gamma-sterilized Charnley Ogee acetabular cup. The Spectron EF was used with either EtO-sterilized non-cross-linked polyethylene (Reflection All-Poly) or highly cross-linked (Reflection All-Poly XLPE) cups, combined with either cobalt chrome (CoCr) or Oxinium femoral heads. The patients were followed with repeated RSA measurements for 2 years. After 2 years, the EtO-sterilized non-cross-linked Reflection All-Poly cups had more than four times higher proximal penetration than its highly cross-linked counterpart. Use of Oxinium femoral heads did not affect penetration at 2 years compared to heads made of CoCr. Further follow-up is needed to evaluate the benefits, if any, of Oxinium femoral heads in the clinical setting. The Charnley Ogee was not outperformed by the more recently introduced implants in our study. We conclude that this prostheses still represents a standard against which new implants can be measured.
机译:这项随机研究的目的是比较150例患者中五种不同的水泥固定全髋关节的磨损和迁移情况。患者接受的是22.2 mm头的Charnley股骨柄或28 mm头的Spectron EF股骨柄。 Charnley与经伽玛消毒的Charnley Ogee髋臼杯连接。 Spectron EF与EtO灭菌的非交联聚乙烯(Reflection All-Poly)或高度交联(Reflection All-Poly XLPE)杯一起使用,与钴铬(CoCr)或Oxinium股骨头结合使用。对患者进行了2年的重复RSA测量。 2年后,经过EtO消毒的非交联反射式全聚杯比其高度交联的反射杯具有更高的近端穿透力四倍以上。与CoCr制成的头相比,使用Oxinium股骨头在2年内不会影响穿透。需要进一步随访以评估Oxinium股骨头在临床环境中的益处(如果有)。 Charnley Ogee在我们的研究中并没有比最近引入的植入物好。我们得出的结论是,这种假体仍然是可以测量新植入物的标准。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号