Effects of covered solid sorbent tube sample holders on organic vapor measurements.

机译:覆盖的固体吸附剂管样品架对有机蒸气测量的影响。

获取原文
获取原文并翻译 | 示例

摘要

A study was conducted to examine whether there are significant differences between organic vapor concentrations measured using charcoal tubes with three different configurations: uncovered sample holder (open tube), SKC, and Buck brand covered sample holders. A fractional factorial experimental design was used with the following factors and levels: vapor (n-hexane vs. m-xylene), pump type (pulsating vs. continuous), exposure profile (variable vs. constant), flow rate (30?mL/min vs. 200?mL/min), duration (30?min vs. 80?min), and sample placement (mannequin vs. free hanging). Two of each sampler configuration (six total) were placed in an exposure chamber, and a dynamic test-atmosphere generation system was used to prepare atmospheres containing approximately 12-15?ppm n-hexane or m-xylene with exposure profiles and sampling conducted according to a run sheet generated for the experimental design. A total of 24 runs were completed with six samplers per run, yielding 144 samples that were analyzed by gas chromatography/flame ionization detector. Concentration results for each pair of SKC and Buck covered sample holders were averaged and normalized by dividing by the average result for the open tube sampler from the same run to eliminate the effect of daily variation in chamber concentrations. The resulting ratio of covered sample tube holder and open tube concentrations was used as the response variable. Results of analysis of variance using the general linear model (MINITAB 16) identified statistically significant main effects and/or interactions for pump type, exposure profile, flow rate, and sample holder. However, the magnitude of the effects was generally less than 10%, and overall mean concentration ratios were 0.989 and 1.02 for the Buck and SKC sample holders, respectively. These results show good agreement between covered sample holder results and open tube measurements and demonstrate that exposure assessment errors resulting from the use of covered sorbent tube sample holders for organic vapor monitoring are relatively small (<10%) and not likely to be of practical importance.
机译:进行了一项研究,以检查使用具有三种不同配置的木炭管测得的有机蒸气浓度之间是否存在显着差异:未覆盖的样品架(开口管),SKC和带有Buck品牌的覆盖样品架。使用分数阶乘实验设计,并具有以下因子和水平:蒸气(正己烷与间二甲苯),泵类型(脉冲与连续),暴露曲线(变量与常数),流速(30?mL) / min与200?mL / min),持续时间(30?min与80?min)和样品放置(人体模型与自由悬挂)。每个采样器配置中的两个(总共六个)放置在曝光室中,并使用动态测试大气生成系统来制备包含约12-15ppm ppm正己烷或间二甲苯的大气,并进行曝光,并根据到为实验设计生成的运行表。共完成24个运行,每个运行有六个采样器,得到144个样品,并通过气相色谱/火焰电离检测器进行了分析。将每对SKC和Buck覆盖的样品架的浓度结果取平均值,并通过除以同一操作的开管采样器的平均结果进行归一化,以消除腔室浓度每日变化的影响。所覆盖的样品管支架和开口管浓度的比率用作响应变量。使用通用线性模型(MINITAB 16)进行方差分析的结果确定了泵类型,暴露曲线,流速和样品架具有统计学意义的主要影响和/或相互作用。但是,这种影响的幅度通常小于10%,Buck和SKC样品架的总体平均浓度比分别为0.989和1.02。这些结果表明,带盖的样品架结果与敞口管测量结果之间具有良好的一致性,并表明由于使用带盖的吸附剂管样品架进行有机蒸气监测而导致的暴露评估误差相对较小(<10%),不太可能具有实际意义。 。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号