A New Spray Device to Deliver Topical Ocular Medication: Penetration of Fluorescein to the Anterior Segment

机译:一种新型的局部眼药喷雾装置:荧光素渗透到前段

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To measure the penetration of fluorescein into the anterior ocular compartments after exposure of the cornea to a mist of aerosol droplets. Methods: This was an open-label proof-of-principle trial. Eighteen healthy volunteers were asked to participate. A conventional (50L) drop of fluorescein solution (20mg/mL) was administered to the right eye; an ocular mist (10L) of the same solution was applied to the left eye. Autofluorescence (photons/s) was measured in the cornea, the anterior chamber (AC), and the lens before administration and at 1, 2, 5, 10, 20, 50, and 100min thereafter. The area under the curve (AUC) was calculated. For the vitreous cavity, measurements were performed at baseline and after 100min. Results: All participants completed the study. AUC (meanSD) for the cornea was (363 +/- 431)x10(4) photons after drop application and (154 +/- 265)x10(4) photons after the mist (P=0.005). For the AC, these values were (6.9 +/- 10.3)x10(4) and (2.9 +/- 5.4)x10(4) photons, respectively (P=0.14). Autofluorescence data obtained in the lens did not allow reliable AUC calculations. Autofluorescence in the vitreous at 100min did not significantly exceed the level at baseline. Conclusion: It was demonstrated that fluorescein applied to the ocular surface with the spray device enters the AC. The total amount of fluorescein molecules reaching the ocular surface by the 2 methods of administration, however, is not equivalent. Therefore, no definitive conclusions on relative bioavailability can be drawn from this experiment.
机译:目的:测量在角膜暴露于气雾滴雾后荧光素向眼前房的渗透。方法:这是一项开放性原则验证试验。 18名健康志愿者被要求参加。常规(50L)滴荧光素溶液(20mg / mL)施用于右眼;将相同溶液的眼雾(10L)涂在左眼上。在给药前以及给药后1、2、5、10、20、50和100分钟,在角膜,前房(AC)和晶状体中测量自体荧光(光子/秒)。计算曲线下面积(AUC)。对于玻璃体腔,在基线和100分钟后进行测量。结果:所有参与者均完成了研究。滴眼后角膜的AUC(meanSD)为(363 +/- 431)x10(4)光子,雾后为(154 +/- 265)x10(4)光子(P = 0.005)。对于AC,这些值分别是(6.9 +/- 10.3)x10(4)和(2.9 +/- 5.4)x10(4)光子(P = 0.14)。在镜片中获得的自发荧光数据无法进行可靠的AUC计算。玻璃体在100分钟时的自发荧光未明显超过基线水平。结论:已经证明,用喷雾装置涂在眼表上的荧光素进入AC。然而,通过两种给药方法到达眼表面的荧光素分子的总量并不相等。因此,无法从该实验得出相对生物利用度的明确结论。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号