首页> 外文期刊>Journal of neurointerventional surgery >Reducing radiation dose while maintaining diagnostic image quality of cerebral three-dimensional digital subtraction angiography: An in vivo study in swine
【24h】

Reducing radiation dose while maintaining diagnostic image quality of cerebral three-dimensional digital subtraction angiography: An in vivo study in swine

机译:降低辐射剂量,同时保持脑三维数字减影血管造影的诊断图像质量:猪的体内研究

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Three-dimensional digital subtraction angiography (3D-DSA) is a modern technique that allows for better appreciation of complex vascular lesions. This study evaluates the impact of various dose reduction strategies on 3D-DSA image quality. Methods: The standard manufacturer 5 s 0.36 μGy/frame setting was modified to create lower dose 3D-DSA protocols by varying the acquisition time (5 or 3 s) and/or dose per frame (0.36, 0.24, 0.17, and 0.10 μGy/f). All protocols were evaluated in three swine. Four raters measured a segment of the external carotid artery on two-dimensional multiplanar reconstruction images. The raters were also presented with three-dimensional volume rendered images from all protocols in a blinded manner and asked to choose the superior image. A full model analysis of variance with repeated measure factors was performed to compare mean differences in measurements between protocols. Results: Measurement differences between the standard and low dose protocols were not clinically significant (<0.5 mm). All raters demonstrated high inter-rater reliability. The 5 s protocols were considered as qualitatively superior to the 3 s protocols. Delivered system doses ranged from 43.8 to 6.5 mGy. The 5 s 0.10 μGy/frame protocols generated 65-68% less delivered dose compared with the 5 s 0.36 μ Gy/frame setting. Conclusions: Low dose 3D-DSA protocols with preserved image quality are achievable, and can help reduce unnecessary radiation exposure to both patients and operators. The 5 s low dose protocols generated clinically acceptable and superior images compared with the 3 s protocols, suggesting a more important role for acquisition time than dose per frame to maintain image quality.
机译:背景:三维数字减影血管造影(3D-DSA)是一项现代技术,可让您更好地欣赏复杂的血管病变。这项研究评估了各种减量策略对3D-DSA图像质量的影响。方法:修改标准制造商的5 s 0.36μGy/帧设置,以通过更改采集时间(5或3 s)和/或每帧剂量(0.36、0.24、0.17和0.10μGy/来创建更低剂量的3D-DSA协议。 F)。在三头猪中评估所有方案。四个评估者在二维多平面重建图像上测量了颈外动脉的一部分。还以不知情的方式为评估者提供了来自所有协议的三维立体渲染图像,并要求他们选择上级图像。对具有重复测量因子的方差进行完整模型分析,以比较方案之间的测量均值差异。结果:标准和低剂量方案之间的测量差异在临床上不显着(<0.5 mm)。所有评估者都显示出较高的评估者间可靠性。 5s协议被认为在质量上优于3s协议。输送的系统剂量范围为43.8至6.5 mG​​y。与5 s 0.36μGy /帧设置相比,5 s 0.10μGy/帧协议产生的递送剂量减少了65-68%。结论:可以实现低剂量的3D-DSA协议,并保持图像质量,并且可以帮助减少对患者和操作者的不必要辐射。与3 s方案相比,5 s的低剂量方案产生了临床上可接受的优质图像,这表明采集时间比每帧剂量更重要,以保持图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号