MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy

机译:经尿道超声在3.0T下对人类前列腺进行MR测温

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and Methods: Eleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. Results: For all scan parameters investigated, the temperature standard deviation within the prostate was less than 1°C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. Conclusion: A good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices.
机译:目的:研究分段梯度回波回波平面成像(EPI)技术在人前列腺中3 T时的质子共振频率(PRF)位移温度计的空间,时间和温度分辨率,并确定适当的序列磁共振成像(MRI)控制的经尿道超声热疗法的参数。材料和方法:11名健康志愿者(年龄范围23-58)在3 T下用16通道躯干线圈进行扫描,以研究梯度回波EPI测温序列的行为。评估了11种不同参数集的温度稳定性和几何变形。在另外五名志愿者中,测量了前列腺T2 *。结果:对于所研究的所有扫描参数,前列腺内的温度标准偏差均小于1°C,而畸变小于1mm。 TE值较高(最高25毫秒),体素尺寸较大和EPI因子较低时,温度稳定性最佳,但这必须与良好的空间和时间分辨率的要求相平衡。前列腺T2 *值范围为30-50毫秒。结论:在TE = 15毫秒,体素大小= 1.14毫米,EPI因子= 9的情况下,在温度稳定性和时空分辨率之间取得了良好的平衡,这对九个切片的动态扫描时间为7.2秒。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号