It has been shown previously that a significant reduction of the acoustic noise in standard, slow magnetic resonance imaging (MRI) sequences is achieved when the gradient pulses have long sinusoidal ramps. An improvement of this method is now presented for fast gradient echo (FLASH) and fast spin echo (RARE) sequences. The new strategy consists of using a sinusoidal readout pulse with no plateau and extending the phase encoding pulses to the entire readout period. The distribution of k-space samples is no longer equidistant, and images have to be reconstructed with gridding. A reduction of the acoustic noise by 20-40 dBA was obtained with respect to standard sequences installed by the scanner manufacturer, which is 10-20 dBA better than that with the previously proposed method of sinusoidal ramps. J. Magn. Reson. Imaging 2001;13:960-966. Copyright 2001 Wiley-Liss, Inc.
机译:先前已经表明,当梯度脉冲具有长正弦波斜坡时,可以在标准的慢速磁共振成像(MRI)序列中显着降低声噪声。现在,针对快速梯度回波(FLASH)和快速自旋回波(RARE)序列提出了该方法的改进。新策略包括使用无平稳状态的正弦读数脉冲,并将相位编码脉冲扩展到整个读数周期。 k空间样本的分布不再等距,必须使用网格重建图像。相对于扫描仪制造商安装的标准序列,声音噪声降低了20-40 dBA,这比以前提出的正弦波斜坡方法要好10-20 dBA。 J.Magn。雷森成像2001; 13:960-966。版权所有2001 Wiley-Liss,Inc.