首页> 外文期刊>Journal of magnetic resonance imaging: JMRI >Measurements of occupational exposure to switched gradient and spatially-varying magnetic fields in areas adjacent to 1.5 T clinical MRI systems.
【24h】

Measurements of occupational exposure to switched gradient and spatially-varying magnetic fields in areas adjacent to 1.5 T clinical MRI systems.

机译:在与1.5 T临床MRI系统相邻的区域中,职业对暴露于转换梯度和空间变化磁场的暴露的测量。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To determine if magnetic field exposure close to two clinical 1.5 T magnetic resonance imaging (MRI) scanners during image acquisition and when moving in the spatially-varying static magnetic field is compliant with European Union (EU) Directive 2004/40/EC (the Directive). MATERIALS AND METHODS: Using commercially available equipment we measured the magnetic flux density around the scanners during two clinical pulse sequences. The data was compared with frequency-dependent limits that will limit occupational exposure following transposition of the Directive into national law in 2008. The static magnetic field was measured around the scanners and the exposure from movement within this field was simulated. RESULTS: The whole-body exposure experienced when standing close to the face of the magnet exceeds the limits in the Directive on the two scanners tested during clinical sequences. Simulation of movement toward the scanner shows that speeds must be restricted to 1/5 of normal walking speed to comply with the Directive. CONCLUSION: EU Directive 2004/40/EC will have a major impact on the current use and future development of MRI due to limitations on exposure to time-varying gradient fields and movement within the spatially-varying static field. This will make interventional work impossible and routine MRI use impracticable in Europe.
机译:目的:确定在图像采集期间以及在空间变化的静磁场中移动时,磁场暴露是否接近两台临床1.5 T磁共振成像(MRI)扫描仪是否符合欧盟(EU)指令2004/40 / EC(指令)。材料与方法:我们使用市售设备在两个临床脉冲序列期间测量了扫描仪周围的磁通密度。将数据与频率相关的限制进行了比较,该限制将限制该指令在2008年转化为国家法律后的职业暴露。在扫描仪周围测量了静磁场,并模拟了该磁场在运动范围内的暴露。结果:当靠近磁体表面站立时全身经历的暴露超过了在临床程序中测试的两台扫描仪上的指令限制。朝扫描仪运动的模拟表明,为了符合指令,必须将速度限制为正常行走速度的1/5。结论:由于暴露于随时间变化的梯度场和在空间变化的静态范围内移动,欧盟指令2004/40 / EC将对MRI的当前使用和未来发展产生重大影响。这将使介入工作变得不可能,并且在欧洲无法进行常规的MRI检查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号