首页> 外文期刊>Journal of interventional cardiac electrophysiology: an international journal of arrhythmias and pacing >Rapid acquisition of high-resolution electroanatomical maps using a novel multielectrode mapping system.
【24h】

Rapid acquisition of high-resolution electroanatomical maps using a novel multielectrode mapping system.

机译:使用新型多电极测绘系统快速获取高分辨率的电解剖图。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Conventional electroanatomical mapping systems employ roving catheters with one or a small number of electrodes. Maps acquired using these systems usually contain a small number of points and take a long time to acquire. Use of a multielectrode catheter could facilitate rapid acquisition of higher-resolution maps through simultaneous collection of data from multiple points in space; however, a large multielectrode array could potentially limit catheter maneuverability. The purpose of this study was to test the feasibility of using a novel, multielectrode catheter to map the right atrium and the left ventricle.Electroanatomical mapping of the right atrium and the left ventricle during both sinus and paced rhythm were performed in five swine using a conventional mapping catheter and a novel, multielectrode catheter.Average map acquisition times for the multielectrode catheter (with continuous data collection) ranged from 5.2 to 9.5 min. These maps contained an average of 2,753 to 3,566 points. Manual data collection with the multielectrode catheter was less rapid (average map completion in 11.4 to 18.1 min with an average of 870 to 1,038 points per map), but the conventional catheter was slower still (average map completion in 28.6 to 32.2 min with an average 120 to 148 points per map).Use of this multielectrode catheter is feasible for mapping the left ventricle as well as the right atrium. The multielectrode catheter facilitates acquisition of electroanatomical data more rapidly than a conventional mapping catheter. This results in shorter map acquisition times and higher-density electroanatomical maps in these chambers.
机译:常规的电解剖标测系统采用带有一个或少量电极的粗纱导管。使用这些系统获取的地图通常包含少量点,并且需要很长时间才能获取。通过同时收集来自多个空间点的数据,多电极导管的使用可以促进快速获取更高分辨率的地图。然而,大型的多电极阵列可能会限制导管的可操作性。这项研究的目的是测试使用新颖的多电极导管绘制右心房和左心室的可行性。在5头猪中,使用窦道和节律性节奏对右心房和左心室的电解剖图常规测绘导管和新颖的多电极导管。多电极导管(连续采集数据)的平均地图采集时间为5.2至9.5分钟。这些地图平均包含2753至3566点。用多电极导管进行手动数据收集的速度较慢(平均地图完成时间为11.4至18.1分钟,每个地图平均870至1,038点),但常规导管的速度仍然较慢(平均地图完成时间为28.6至32.2分钟,平均每个地图120到148个点)。使用此多电极导管可用于绘制左心室和右心房。与传统的标测导管相比,多电极导管有助于更快地获取电解剖数据。这样可以缩短这些腔室中的地图获取时间,并获得更高密度的电解剖图。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号