Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes.

机译:用动态轮廓眼压计,眼反应分析仪和戈德曼眼压计在青光眼中测量中央角膜厚度和角膜滞后对眼压的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The dynamic contour tonometer (DCT; Pascal tonometer) and the Ocular Response Analyzer (ORA) are novel tonometers designed to measure intraocular pressure (IOP) independent of corneal properties and central corneal thickness (CCT), respectively. We wanted to compare the corneal compensated IOP (IOPcc) as measured by ORA with IOP values measured by DCT and Goldmann applanation tonometry (GAT) with respect to IOP readings and the influence of corneal hysteresis (CH) and CCT in glaucoma patients. PATIENTS AND METHODS: In a study of 94 glaucomatous eyes, IOP measurements by ORA, DCT, and GAT were compared, and the effects of CCT and CH were analyzed. All measurements were taken by 1 of the authors only. RESULTS: The Mean CCT was 550+/-44 mum and mean CH 10.24+/-3.3 mm Hg. The mean value for IOPcc was 17.94+/-5.9 mm Hg, 15.14+/-3.7 mm Hg for DCT, and 14.3+/-4.3 mm Hg for GAT. The mean difference was 3.65+/-3.85 mm Hg between IOPcc and GAT and 2.80+/-4.9 mm Hg between IOPcc and DCT (P<0.001) and -0.85+/-3.3 mm Hg between GAT and DCT (P=0.015). None of the methods was related to CCT. In contrast, CH was highly statistically significant related to IOPcc (P<0.0001), whereas GAT and DCT showed only weak relation to CH (P=0.05). CONCLUSIONS: IOP measurements with the ORA are significantly higher than DCT or GAT with no apparent linear correction factor. To date, we cannot estimate the impact of differences in CH in relation to IOP and its measurement.
机译:目的:动态轮廓眼压计(DCT; Pascal眼压计)和眼响应分析仪(ORA)是新颖的眼压计,旨在分别测量独立于角膜特性和中央角膜厚度(CCT)的眼内压(IOP)。我们希望将通过ORA测量的角膜补偿IOP(IOPcc)与通过DCT和Goldmann压平眼压计(GAT)测量的IOP值相对于IOP读数以及青光眼患者的角膜滞后(CH)和CCT的影响进行比较。病人与方法:在94只青光眼的研究中,比较了ORA,DCT和GAT对眼压的测量结果,并分析了CCT和CH的影响。所有测量仅由1位作者进行。结果:平均CCT为550 +/- 44毫米,平均CH为10.24 +/- 3.3毫米汞柱。 IOPcc的平均值为17.94 +/- 5.9 mm Hg,对于DCT为15.14 +/- 3.7 mm Hg,对于GAT为14.3 +/- 4.3 mm Hg。 IOPcc和GAT之间的平均差异为3.65 +/- 3.85 mm Hg,IOPcc和DCT之间的平均差异为2.80 +/- 4.9 mm Hg(P <0.001),GAT和DCT之间的-0.85 +/- 3.3 mm Hg(P = 0.015) 。没有一种方法与CCT有关。相比之下,CH与IOPcc相关性在统计学上高度显着(P <0.0001),而GAT和DCT仅与CHP相关性较弱(P = 0.05)。结论:ORA的眼压测量值明显高于DCT或GAT,无明显的线性校正因子。迄今为止,我们无法估计CH的差异对IOP及其度量的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号