【24h】

Intradevice and Interdevice Agreement Between a Rebound Tonometer, Icare PRO, and the Tonopen XL and Kowa Hand-held Applanation Tonometer When Used in the Sitting and Supine Position

机译:坐姿和仰卧位使用时,回弹眼压计,Icare PRO,Tonopen XL和Kowa手持压平眼压计之间的设备内和设备间协议

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose:The aim of the study was to investigate the agreement between a new portable tonometer, Icare PRO, and the Tonopen XL and Kowa hand-held applanation tonometers (HAT).Methods:The right eyes of 127 healthy subjects were enrolled. Intraocular pressure (IOP) was measured in both sitting and supine positions using the Icare PRO, Tonopen XL, and Kowa HAT tonometers. The repeatability of the IOP measurements was evaluated by calculating intraclass correlation coefficients. Between-method agreements of tonometer measurements were evaluated using Bland-Altman analysis.Results:Intradevice agreement: The intraclass correlation coefficients (sitting, supine) of Icare PRO, Tonopen XL, and Kowa HAT were (0.863, 0.656), (0.845, 0.819), and (0.957, 0.956), respectively.Interdevice agreement: The Bland-Altman analysis revealed that, in the sitting position, the mean differences between Icare PRO and Tonopen XL, and between Icare PRO and Kowa HAT were -0.43 and 0.43 mm Hg, respectively (95% limits of agreement: -6.24 to 5.34 mm Hg, -4.04 to 4.90 mm Hg). In the supine position, the corresponding mean differences were -0.88 and 0.14 mm Hg (95% limits of agreement: -5.66 to 3.91 mm Hg, -4.06 to 4.33 mm Hg). IOP differences between Icare PRO and the other tonometers were unaffected by central corneal thickness.Conclusions:The repeatability of Icare PRO was slightly lower in the supine position than in the sitting position. Although Icare PRO underestimated IOP values in eyes with higher IOP when compared with Tonopen XL and Kowa HAT in both positions, we observed good interdevice agreement between Icare PRO and both Tonopen XL and Kowa HAT.
机译:目的:本研究的目的是调查新型便携式眼压计Icare PRO与Tonopen XL和Kowa手持式压平眼压计(HAT)之间的一致性。方法:招募了127名健康受试者的右眼。使用Icare PRO,Tonopen XL和Kowa HAT眼压计测量坐位和仰卧位的眼压(IOP)。通过计算类内相关系数来评估IOP测量的可重复性。结果:设备内部一致性:Icare PRO,Tonopen XL和Kowa HAT的类内相关系数(坐位,仰卧位)分别为(0.863、0.656),(0.845、0.819)设备间一致性:Bland-Altman分析显示,在坐姿下,Icare PRO和Tonopen XL之间,Icare PRO和Kowa HAT之间的平均差分别为-0.43和0.43 mm分别为Hg(协议的95%限制:-6.24至5.34毫米汞柱,-4.04至4.90毫米汞柱)。在仰卧位置,相应的平均差异为-0.88和0.14 mm Hg(95%的一致限:-5.66至3.91 mm Hg,-4.06至4.33 mm Hg)。 Icare PRO与其他眼压计的IOP差异不受中央角膜厚度的影响。结论:仰卧位Icare PRO的可重复性略低于坐位。尽管与两个位置的Tonopen XL和Kowa HAT相比,Icare PRO低估了具有较高IOP的眼睛的IOP值,但我们观察到Icare PRO与Tonopen XL和Kowa HAT的设备间一致性良好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号