Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer

机译:新型反弹式眼压计Icare PRO和Icare ONE与Goldmann眼压计的临床评价

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose:To compare the participant-obtained intraocular pressures (IOPs) using Icare ONE and the clinician-obtained values using Icare PRO, both rebound tonometers, with Goldmann tonometry (GAT) values and analyze the ease of use of Icare ONE.Methods:One hundred fifty participants were included (60 normal controls, 90 patients with glaucoma/ocular hypertension). The participants measured the IOP 3 times using Icare ONE; a clinician measured the IOP once using Icare PRO. The instruments were used randomly. Clinical data were evaluated to analyze the difficulty of the technique, the effect on the results, and the ease of use of Icare ONE.Results:The mean IOPs with GAT, Icare ONE, and Icare PRO were 16.64.43, 17.5 +/- 5.42, and 16.6 +/- 4.77 mm Hg, respectively. The participant-measured IOP values were within +3 mm Hg of the GAT values in 67.1% of eyes with Icare ONE and in 79.6% with Icare PRO. The limits of agreement were higher with Icare ONE compared with Icare PRO. IOP value errors were found with Icare ONE in eyes with low and high GAT-IOP. The areas under the curve to detect IOPs of 21 mm Hg or higher (GAT) exceeded 0.80 with both tonometers. Young participants reported better ease of use with Icare ONE. No other factors were related to the results.Conclusions:Icare ONE may be useful for patients monitoring their IOP values; most individuals can use the device after a short training session. Icare PRO had better results compared with Icare ONE in all IOP ranges.
机译:目的:比较使用Icare ONE的参与者获得的眼内压(IOP)和使用Icare PRO的临床医生获得的眼压(两个回弹眼压计)和Goldmann眼压(GAT)值,并分析Icare ONE的易用性。包括150名参与者(60名正常对照,90名青光眼/高眼压患者)。参与者使用Icare ONE测量了3次IOP;临床医生使用Icare PRO测量一次IOP。这些仪器是随机使用的。评估临床数据以分析该技术的难度,对结果的影响以及Icare ONE的易用性。结果:GAT,Icare ONE和Icare PRO的平均IOP为16.64.43,17.5 +/-分别为5.42和16.6 +/- 4.77毫米汞柱。参与者测得的IOP值在使用Icare ONE的67.1%的眼睛和使用Icare PRO的79.6%的眼睛的GAT值的+3 mm Hg之内。与Icare PRO相比,Icare ONE的协议限制更高。使用Icare ONE在低和高GAT-IOP的眼睛中发现IOP值错误。两种眼压计的曲线下面积均可用于检测21 mm Hg或更高的眼压(GAT),均超过0.80。年轻的参与者报告说,使用Icare ONE更好的易用性。结论:Icare ONE可能对监测其IOP值的患者有用;短期培训后,大多数人都可以使用该设备。在所有IOP范围内,与Icare ONE相比,Icare PRO具有更好的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号