首页> 外文期刊>Journal of forensic sciences. >Organic explosives analysis using on column-ion trap EI/NICI GC-MS with an external source.
【24h】

Organic explosives analysis using on column-ion trap EI/NICI GC-MS with an external source.

机译:使用带有外部源的柱离子阱EI / NICI GC-MS分析有机爆炸物。

获取原文
获取原文并翻译 | 示例
           

摘要

In this study, a standard method by gas chromatography coupled with mass spectrometry (GC/MS) for the analysis of classical organic explosives was developed. This method was validated in the EI mode, based on the XPT 90-210 standard method. Detection limits (LOD) and quantitation limits (LOQ) were both determined using electronic impact (EI) and negative ion chemical ionization (NICI) modes. These were compared and results showed that in the NICI mode, detection limits were lower than in the EI mode, thus NICI mode appeared to be the best way to analyze nitrate esters. Results of ion trap MS detection were then compared with those obtained in a previous study with single quadrupolar technology. Major ions that were obtained using ion trap MS detection in these two modes were reported.
机译:在这项研究中,开发了一种气相色谱-质谱联用(GC / MS)分析经典有机炸药的标准方法。基于XPT 90-210标准方法,该方法在EI模式下得到了验证。检测限(LOD)和定量限(LOQ)均使用电子碰撞(EI)和负离子化学电离(NICI)模式确定。比较了这些结果,结果表明,在NICI模式下,检测限低于EI模式,因此NICI模式似乎是分析硝酸酯的最佳方法。然后将离子阱质谱检测的结果与先前使用单四极杆技术的研究结果进行了比较。报告了使用离子阱质谱检测器在这两种模式下获得的主要离子。