首页> 外文期刊>Journal of Endodontics: Official Journal of American Association of Endodontists >Clinical accuracy of a new apex locator with an automatic compensation circuit.
【24h】

Clinical accuracy of a new apex locator with an automatic compensation circuit.

机译:具有自动补偿电路的新型根尖定位器的临床准确性。

获取原文
获取原文并翻译 | 示例
       

摘要

A new circuit was designed to automatically compensate for measurement errors of an electronic apex locator in various electrolytes. Thirty-one root canals were clinically tested for accuracy. A file was inserted into the canal until the apex signal was obtained, at which point the file was immobilized with glass-ionomer cement. After extraction, the apical area was exposed and the position of the file tip was examined under an operating microscope. Distances from the major foramen and cemento dentinal junction (CDJ) were recorded. The average distance from the major foramen was -0.13 mm with a range of -1.28 mm and +0.46 mm. The average distance in 26 detectable CDJ samples was +0.18 mm with a range of -0.98 mm and +0.65 mm. The measurements, which were within +/- 0.5 mm, were 94% (29/31) from the major foramen and 92% (24/26) from the CDJ. Measurement consistencies within one SD were 81% for the major foramen and 65% for the CDJ, respectively. Measurements within two SD were 97% for the major foramen and 92% for the CDJ. There were no differences between the smaller (< # 25) and larger apical foramens (< or = 25) or vital and nonvital pulps.
机译:设计了一种新电路,以自动补偿各种电解质中电子顶点定位器的测量误差。临床检查了31条根管的准确性。将锉刀插入根管直至获得顶点信号,此时将锉刀用玻璃离聚物水泥固定。提取后,将根尖区域暴露出来,并在手术显微镜下检查锉刀尖端的位置。记录距主要孔和牙骨质牙本质连接处(CDJ)的距离。与主要孔的平均距离为-0.13 mm,范围为-1.28 mm和+0.46 mm。 26个可检测CDJ样本中的平均距离为+0.18毫米,范围为-0.98毫米和+0.65毫米。测量值在+/- 0.5毫米范围内,距主要孔的94%(29/31),来自CDJ的92%(24/26)。一个SD内的测量一致性主要孔为81%,CDJ为65%。两个标准差内的测量值分别是主要孔的97%和CDJ的92%。较小的(<#25)和较大的顶端甲孔(<或= 25)或重要和非重要牙髓之间没有差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号