Evaluating the Effects of Different Dental Devices on Implantable Cardioverter Defibrillators

机译:评估不同的牙科设备对植入式心脏复律除颤器的影响

获取原文
获取原文并翻译 | 示例
       

摘要

Introduction: The implantable cardioverter defibrillator (ICD) is an electronic device that emits electrical signals to the heart via lead wires and electrodes. It is used for cardiac rhythm monitoring and treatment. Because electronic dental devices have been shown to produce electromagnetic fields, we hypothesize that they may interfere with ICD function. Methods: Nine dental devices (heat carrier, electronic apex locator, electric pulp tester, unipolar electrosurgery unit, electric motor, curing light, and 3 gutta-percha guns) were tested in this study for their ability to interfere with the function of 4 ICDs (2 single-chambered and 2 dual-chambered ICDs). ICD activity was monitored for 30 seconds using an ICD programmer (Medtronic 2090; Minneapolis, MN) and evaluated through an electrogram test strip printout. Results: Electromagnetic interference was detected with the electric motor, curing light, electric pulp tester, and electrosurgery unit although no electromagnetic disturbances were detected with these devices. No electrornagnetic interferences were observed for the gutta-percha guns, heat carrier, and apex locator. However, the electrosurgery unit affected the dual-chambered ICD (Consulta CRT-D, Medtronic) and delivered therapies for fibrillation when no ventricular fibrillation was present. Conclusions: Our results suggest that the electrosurgery unit produces electromagnetic disturbances with unwanted therapy delivery shock and potentially clinically significant outcomes.
机译:简介:植入式心脏复律除颤器(ICD)是一种电子设备,可通过导线和电极向心脏发出电信号。它用于心律监测和治疗。由于已经证明电子牙科设备会产生电磁场,因此我们假设它们可能会干扰ICD功能。方法:本研究测试了九种牙科设备(热载体,电子根尖定位器,电浆测试仪,单极电外科设备,电动机,固化灯和3支牙胶枪)干扰4种ICD功能的能力。 (2个单腔ICD和2个双腔ICD)。使用ICD编程器(Medtronic 2090;明尼阿波利斯,明尼苏达州)对ICD活性进行30秒钟的监测,并通过电描记条测试输出进行评估。结果:尽管未检测到电磁干扰,但使用电动机,固化灯,电动纸浆测试仪和电外科设备检测到电磁干扰。对于古塔胶枪,热载体和顶点定位器,未观察到电磁干扰。但是,电外科单元影响了双腔ICD(Consulta CRT-D,Medtronic),并且在不存在心室纤颤的情况下提供了纤颤疗法。结论:我们的结果表明,电外科单元会产生电磁干扰,并产生不必要的治疗电击,并可能在临床上产生重大后果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号