Effects of two nickel-titanium instrument systems, Mtwo versus ProTaper universal, on root canal geometry assessed by micro-computed tomography.

机译:Mtwo与ProTaper Universal这两种镍钛仪器系统对通过微计算机断层扫描评估的根管几何形状的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

INTRODUCTION: The aim of this study was to compare the effects of Mtwo and ProTaper Universal (PTU) on root canal geometry three-dimensionally by using micro-computed tomography. METHODS: A total of 20 canals were prepared with Mtwo or PTU system. Micro-computed tomography scanner was used to scan the specimens before and after instrumentation. Three-dimensional images were reconstructed. The parameters evaluated included canal surface area, volume, structure model index, thickness, straightening, canal transportation, and uninstrumented surface area. RESULTS: Preparation significantly changed canal surface area, volume, structure model index, thickness, and canal curvature. There were no significant differences between instrument types concerning these parameters and uninstrumented surface area. The canals prepared with PTU showed larger values of transportation compared with those in Mtwo group at the apical third. CONCLUSIONS: Both of the instrumentation systems produced canal preparations with adequate geometry. PTU produced larger transportation at apical third.
机译:引言:本研究的目的是通过使用微计算机断层扫描技术来比较Mtwo和ProTaper Universal(PTU)对三维根管几何学的影响。方法:用Mtwo或PTU系统准备了总共20条运河。在仪器化之前和之后,使用微型计算机断层扫描仪扫描样品。三维图像被重建。评估的参数包括运河表面积,体积,结构模型指数,厚度,矫直,运河运输和非仪表表面积。结果:准备工作显着改变了根管的表面积,体积,结构模型指数,厚度和根管曲率。关于这些参数的仪器类型与未仪器化的表面积之间没有显着差异。与Mtwo组在心尖的第三根相比,用PTU制备的运河具有更大的运输价值。结论:这两种器械系统均生产了具有适当几何形状的牙根管。 PTU在第三位产生了更大的运输量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号