Purpose: To describe technical challenges encountered using the Ovation endograft for abdominal aortic aneurysms and suggest tips and maneuvers for successful management. Technique: Deployment of the Ovation's unsupported main body is often associated with the anteroposterior arrangement of the limb gates instead of the usual side-by-side configuration, rendering contralateral catheterization challenging. Successful catheterization of the contralateral limb can be confirmed by lateral deviation of the ipsilateral stiff guidewire caused by balloon inflation within the contralateral limb. Moreover, failure to cannulate the contralateral limb gate due to persistent impingement of the guidewire or compromise of the inflated rings can be overcome using the transbrachial antegrade approach. Conclusion: Awareness of certain aspects associated with the use of the Ovation device and adoption of the adjunctive techniques herein described helps the interventionist avoid unexpected challenges or manage technical difficulties to maximize the device's safety and effectiveness.
机译:目的:描述使用Ovation内移植物治疗腹主动脉瘤时遇到的技术挑战,并为成功治疗提供建议和技巧。技术:Ovation的无支撑主体的展开通常与肢体闸门的前后排列相关联,而不是通常的并排配置,这使得对侧导管插入具有挑战性。对侧肢体成功导尿可以通过对侧肢体内球囊膨胀引起的同侧刚性导丝的侧向偏移来确认。此外,可以使用经臂顺行方法克服由于导丝的持续撞击或充气环受损而导致无法插入对侧肢体门的可能性。结论:意识到与使用Ovation设备相关的某些方面以及采用本文所述的辅助技术有助于干预人员避免意外的挑战或管理技术难题,以最大程度地提高设备的安全性和有效性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号