Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay

机译:在JCA-BM6010 / C上评估酶促BM测试HbA1c并与Bio-Rad Variant II Turbo,Tosoh HLC 723 G8和AutoLab免疫比浊法进行比较

获取原文
获取原文并翻译 | 示例
           

摘要

Background: A novel enzymatic HbA1c assay was introduced for use in an automated chemistry analyzer. With this unique method, HbA1c and plasma glucose can be measured from the same EDTA tube. We evaluated the analytical performance of this enzymatic HbA1c assay in a JCA-BM6010/C analyzer and compared the HbA1c values with the results from other widely used methodological instruments. Methods: The imprecision, linearity, carry-over and concordance rate of the enzymatic HbA1c test (BM Test HbA1c) using the JCA-BM6010/C analyzer were evaluated. Three hundred and seventy-seven specimens with HbA1c concentrations from 16 to 133 mmol/mol were used for a comparison study with two high performance liquid chromatography methods: Variant II Turbo and Tosoh HLC 723 G8 and the AutoLab Hemoglobin A1c immunoturbidimetry reagent using a Hitachi 7600-110. Forty specimens were used for the glucose method comparison. Results: The HbA1c coefficients of variation of the within-run imprecision for low and high levels were 0.6% and 0.4%, respectively. The linearity of the BM Test HbA1c using the JCA-BM6010/C analyzer was excellent in the range between 31 mmol/mol and 143 mmol/mol. The carry-over rate was 0.2%. The relationships between the BM test and the other three methods were 0.916×Tosoh G8+3.644, r=0.986; 0.887×Bio-Rad Variant II+1.896, r=0.972; and 0.941×AutoLab+4.532, r=0.977. The concordance rates using a cut-off of 48 mmol/mol were 91.5% with Tosoh G8, 82.8% with Bio-Rad Variant II, and 91.0% with AutoLab. The simultaneously assayed plasma glucose with HbA 1c was 1.002×Routine plasma glucose+0.625, r=1.000 Conclusions: The enzymatic BM Test HbA1c in the JCA-BM6010/C analyzer showed excellent precision and linearity, and a minimal carry-over rate. The simultaneously assayed plasma glucose analysis showed good performance.
机译:背景:一种新型的酶促HbA1c检测方法被引入自动化学分析仪中。使用这种独特的方法,可以从同一EDTA管中测量HbA1c和血浆葡萄糖。我们在JCA-BM6010 / C分析仪中评估了这种酶促HbA1c测定的分析性能,并将HbA1c值与其他广泛使用的方法学仪器的结果进行了比较。方法:使用JCA-BM6010 / C分析仪评估酶促HbA1c测试(BM测试HbA1c)的不精确度,线性,残留率和一致性率。 HbA1c浓度从16到133 mmol / mol的377个样本用于两种高效液相色谱方法的比较研究:Variant II Turbo和Tosoh HLC 723 G8以及使用Hitachi 7600的AutoLab血红蛋白A1c免疫比浊法试剂-110。 40个样本用于葡萄糖法比较。结果:低水平和高水平运行时不精确度的HbA1c变异系数分别为0.6%和0.4%。使用JCA-BM6010 / C分析仪的BM测试HbA1c的线性在31 mmol / mol至143 mmol / mol的范围内极佳。残留率为0.2%。 BM测试与其他三种方法的关系为0.916×Tosoh G8 + 3.644,r = 0.986; 0.887×Bio-Rad Variant II + 1.896,r = 0.972;和0.941×AutoLab + 4.532,r = 0.977。使用Tosoh G8,使用48 mmol / mol截留值的一致性率为91.5%,使用Bio-Rad Variant II的一致性率为82.8%,使用AutoLab的一致性率为91.0%。用HbA 1c同时测定的血浆葡萄糖为1.002×常规血浆葡萄糖+ 0.625,r = 1.000结论:JCA-BM6010 / C分析仪中的酶促BM测试HbA1c具有优异的精密度和线性,并且携带率极低。同时测定的血浆葡萄糖分析显示出良好的性能。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号