BACKGROUND: The metrological traceability of values for the catalytic concentration of several enzymes assigned to a calibration material has been assured by following the recently published International Standard ISO 18153. METHODS: A traceable value with a measurement uncertainty was assigned for the catalytic concentration of alanine aminotransferase, creatine kinase, gamma-glutamyltransferase and lactate dehydrogenase in two materials from different sources. These are all measurable quantities, with the primary reference measurement procedure described by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) and a primary calibrator giving metrological traceability to the SI unit of measurement. The metrologically traceable calibration was validated by measuring human serum samples using the primary reference measurement procedure and a routine commercial measurement procedure calibrated with the traceable materials. RESULTS: Results showed that the primary reference procedure, selected manufacturers' procedures and the end-user's routine procedure for each enzyme have the same analytical specificity. Four of eight commercial calibrators tested were commutable, whereas the others had a very small difference in absolute terms, indicating that these materials would be useful for calibration. CONCLUSION: The implementation of a reference system for enzyme measurements was demonstrated that assures the traceability of patient results to SI units.
机译:背景:通过遵循最近发布的国际标准ISO 18153,可以确保分配给校准材料的几种酶的催化值的计量可溯性。方法:为丙氨酸转氨酶的催化浓度指定了具有测量不确定性的可追溯值。 ,肌酸激酶,γ-谷氨酰转移酶和乳酸脱氢酶来自不同来源的两种材料。这些都是可测量的量,国际临床化学和检验医学联合会(IFCC)描述了主要参考测量程序,主要校准器为SI测量单位提供了计量溯源。通过使用主要参考测量程序和使用可追溯材料校准的常规商业测量程序测量人血清样本,验证了可计量溯源的校准。结果:结果表明,每种酶的主要参考程序,所选制造商的程序和最终用户的常规程序具有相同的分析特异性。被测试的八种商用校准器中有四种是可互换的,而其他校准器的绝对值差异很小,表明这些材料可用于校准。结论:证明了酶测量参考系统的实施可确保将患者结果追溯至国际单位。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号