An important point when working with automated immunochemical techniques is to find suitable diluents for high-concentration samples. Although current immunometric techniques give a very wide analytical range and in many clinical situations sample dilutions are not necessary (1), to avoid interferences, the matrix of any diluents used should be as close as possible to that of normal sera from patients. Moreover, matrix problems vary from one sample to another and different immunoassays have different matrix problems (2). Diluents supplied by manufacturers in reagent kits usually do not show the same physicochemical properties as those of human serum, probably for legal or economic reasons. This means that clinical biochemists may have to assume that results obtained for high-concentration samples of different analytes may be erroneous or to prepare a suitable diluent to avoid these mistakes.
机译:使用自动化免疫化学技术时,重要的一点是为高浓度样品找到合适的稀释剂。尽管当前的免疫测定技术提供了非常宽的分析范围,并且在许多临床情况下不需要稀释样品(1),但为了避免干扰,所用任何稀释剂的基质应尽可能接近患者的正常血清。此外,基质问题因样品而异,不同的免疫测定法具有不同的基质问题(2)。制造商在试剂盒中提供的稀释剂通常不具有与人血清相同的理化特性,这可能是出于法律或经济方面的原因。这意味着临床生物化学家可能不得不假设针对不同分析物的高浓度样品获得的结果可能是错误的,或者为了避免这些错误而准备了合适的稀释剂。