首页> 外文期刊>Journal of diagnostic medical sonography: JDMS >The Methods and Effects of Transducer Degradation on Image Quality and the Clinical Efficacy of Diagnostic Sonography
【24h】

The Methods and Effects of Transducer Degradation on Image Quality and the Clinical Efficacy of Diagnostic Sonography

机译:换能器降解对图像质量和超声诊断学的临床效果的方法和效果

获取原文
获取原文并翻译 | 示例
           

摘要

In diagnostic ultrasound examinations, transducer "health" is key to diagnostic efficacy. It is known that individual transducer element integrity within an array is central to overall probe performance and over time, with normal use, elements can cease working or lose sensitivity, leading to a potentially negative impact on the clinical efficacy of the ultrasound examination. Investigating this issue, the authors evaluated transducers with selected elements disabled compared to fully functioning arrays, examined how dead elements affected ultrasound beams, acoustic parameters, flow phantom/tissue phantom results as well as human imaging. Results: As few as 2 consecutive dead elements can materially impact the beam profile; four or more can significantly reduce resolution and penetration, increase the noise floor, and cause Doppler peak velocity errors, flow ambiguity and spectral broadening. Tissue phantoms proved to be equivocal in spotting defective elements. Conclusion: array heath is critical to high-quality, efficacious ultrasound studies and the potential for misdiagnosis increases as array elements degrade.
机译:在超声诊断检查中,换能器的“健康”是诊断功效的关键。众所周知,阵列中单个换能器元件的完整性对于整体探头性能至关重要,并且随着时间的流逝,在正常使用情况下,元件可能会停止工作或失去灵敏度,从而可能对超声检查的临床疗效产生负面影响。为了研究这个问题,作者评估了与全功能阵列相比禁用了选定元素的换能器,研究了死元素如何影响超声波束,声学参数,体模/组织体模结果以及人体成像。结果:仅有2个连续的死元素会实质性地影响光束轮廓;四个或更多会大大降低分辨率和穿透力,增加本底噪声,并导致多普勒峰速度误差,流模糊和频谱展宽。组织体模被证明在发现缺陷元素方面是模棱两可的。结论:阵列健康对高质量,有效的超声研究至关重要,并且随着阵列元件的退化,误诊的可能性也会增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号