摘要

The measurement of troponins in blood has rapidly become an accepted method for detecting myocardial damage. Furthermore, recently redefined criteria for myocardial infarction used to classify patients with acute coronary syndrome have been established on the basis of increased serum/plasma concentrations of cardiac troponin I (cTnl) or T (cTnT). Studies have indicated the prognostic importance of increased troponins in various clinical settings. In fact, incorrect quantification of cTnl may occur due to antibody and non-antibody interferences. The presence of interfering antibodies such as heterophilic antibodies and rheumatoid factors can affect immunoassays and produce either falsely elevated or falsely low results. In addition to these factors, non-antibody interferences such as residual fibrin and microparticles may also be a contributing factor to incorrect measurement of cTnl.
机译:血液中肌钙蛋白的测定已迅速成为检测心肌损伤的公认方法。此外,基于增加的心肌肌钙蛋白I(cTnl)或T(cTnT)的血清/血浆浓度,最近建立了用于对急性冠脉综合征患者进行分类的心肌梗死的重新定义标准。研究表明肌钙蛋白升高在各种临床环境中对预后的重要性。实际上,由于抗体和非抗体干扰,可能会发生cTnl定量错误。诸如异源抗体和类风湿因子之类的干扰抗体的存在会影响免疫测定,并产生错误升高或错误降低的结果。除这些因素外,非抗体干扰物(例如残留的纤维蛋白和微粒)也可能是导致cTnl错误测量的因素。