BACKGROUND: Hyperthyroidism in Graves' disease (GD) is often associated with the production of autoantibodies (TRAb) to the thyrotropin receptor (TSHR). Current manual second generation TRAb assays demonstrate high clinical sensitivity, but are labor-intensive and time consuming. Until recently, technical difficulties prevented the availability of an automatic TRAb assay. METHODS: Development of a fast and fully automated TRAb assay on the Elecsys/cobas e electrochemiluminescence immunoassay platform. RESULTS: A labeled thyroid-stimulating human monoclonal TSHR autoantibody (M22) was used in an automated M22-binding inhibition assay. TRAb are detected by their ability to competitively inhibit M22-binding to solubilized porcine TSHR (pTSHR). High clinical sensitivity could be maintained by assembling multiple TSHR binding sites within a soluble oligomeric immunocomplex for improved TRAb binding. Requirement of sufficient on board stability of the delicate TSHR structure in solution for several days was met by pre-complexation of the pTSHR with a capture antibody to its C-terminus in combination with the use of structure-stabilizing chemical chaperones. Total imprecision coefficient of variation (CV) at 1.71 (approximate cut-off) was found to be 11.4%. TRAb results were available within 30 min. CONCLUSIONS: Availability of a fast and automatic TRAb assay offers an attractive alternative to the manual TRAb assays for the differential diagnosis of hyperthyroidism.
机译:背景:格雷夫斯病(GD)中的甲状腺功能亢进通常与促甲状腺激素受体(TSHR)的自身抗体(TRAb)产生有关。当前的手动第二代TRAb分析显示出很高的临床敏感性,但劳动强度大且耗时。直到最近,技术上的困难仍阻碍了自动TRAb检测的可用性。方法:在Elecsys / cobas e电化学发光免疫测定平台上开发了快速,全自动的TRAb测定。结果:标记的刺激甲状腺的人单克隆TSHR自身抗体(M22)被用于自动化的M22结合抑制试验。通过TRAb竞争性抑制M22与增溶猪TSHR(pTSHR)结合的能力来检测TRAb。通过在可溶性寡聚免疫复合物中组装多个TSHR结合位点以改善TRAb结合,可以保持较高的临床敏感性。通过将pTSHR与捕获其C末端的抗体预复合并结合使用结构稳定的化学分子伴侣,可以满足溶液中精致的TSHR结构在几天内具有足够的板上稳定性的要求。发现总的不精确度变异系数(CV)为1.71(近似截止值),为11.4%。 TRAb结果可在30分钟内获得。结论:快速,自动的TRAb测定法的提供为人工诊断甲状腺功能亢进症的人工TRAb测定法提供了一种有吸引力的替代方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号