首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >High performance liquid chromatography equipped with a cathodic detector and column-switching device as a high-throughput method for a phosphatase assay with p-nitrophenyl phosphate
【24h】

High performance liquid chromatography equipped with a cathodic detector and column-switching device as a high-throughput method for a phosphatase assay with p-nitrophenyl phosphate

机译:配有阴极检测器和柱切换装置的高效液相色谱仪,是通过对硝基苯基磷酸酯进行磷酸酶测定的高通量方法

获取原文
获取原文并翻译 | 示例
           

摘要

LC coupled to an electrochemical detector (LC-EC) operating in cathodic mode with a column-switching system realizes a high-throughput detection of p-nitrophenol (NP). The measurement-time for each NP sample was shortened to 20 s, and the successive analyses of 39 samples was completed within 13 min. In the present system, the limits of detection and quantification were 0.15 and 0.20 muM, respectively, and further, up to 25 muM, a linear calibration curve was afforded. Relative standard deviations for standard solutions of 0.20, 1.0, and 25 muM NP were 4.3, 2.0, and 1.1% (n = 5), respectively. Between-run precisions of the analysis of 5.0 and 25 muM NP over 6 days were 4.8 and 1.3%, respectively. A comparison with the commonly used Bessey-Lowry-Brock method indicates that the present LC-EC is useful for the high-throughput assay of acid and alkaline phosphatases in urine and blood samples with a p-nitrophenyl phosphate substrate. (C) 2005 Elsevier B.V. All rights reserved.
机译:LC与通过柱切换系统在阴极模式下运行的电化学检测器(LC-EC)耦合,可实现对硝基苯酚(NP)的高通量检测。每个NP样品的测量时间缩短到20 s,并且在13分钟内完成了39个样品的连续分析。在本系统中,检测限和定量限分别为0.15和0.20μM,此外,最高25μM,可提供线性校准曲线。标准溶液0.20、1.0和25μMNP的相对标准偏差分别为4.3%,2.0%和1.1%(n = 5)。在6天内对5.0和25μMNP进行分析时的批间精密度分别为4.8%和1.3%。与常用的Bessey-Lowry-Brock方法的比较表明,本LC-EC可用于对-磷酸对硝基苯酯底物的尿液和血液样品中酸和碱性磷酸酶的高通量分析。 (C)2005 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号