With continuing developments in computed tomography (CT) technology and its increasing use of CT imaging, the ionizing radiation dose from CT is becoming a major public concern particularly for high-dose applications such as cardiac imaging. We recently proposed a novel interior tomography approach for x-ray dose reduction that is very different from all the previously proposed methods. Our method only uses the projection data for the rays passing through the desired region of interest. This method not only reduces x-ray dose but scatter as well. In this paper, we quantify the reduction in the amount of x-ray dose and scattered radiation that could be achieved using this method. Results indicate that interior tomography may reduce the x-ray dose by 18% to 58% and scatter to the detectors by 19% to 59% as the FOV is reduced from 50 to 8.6 cm.
机译:随着计算机断层摄影(CT)技术的不断发展及其在CT成像中的越来越多的应用,CT的电离辐射剂量正成为主要的公众关注点,尤其是对于大剂量应用(例如心脏成像)。我们最近提出了一种用于减少X射线剂量的新颖内部层析成像方法,该方法与以前提出的所有方法都大不相同。我们的方法仅将投影数据用于穿过所需感兴趣区域的光线。该方法不仅减少了X射线的剂量,而且还减少了散射。在本文中,我们量化了使用此方法可以实现的X射线剂量和散射辐射数量的减少。结果表明,随着FOV从50 cm减小到8.6 cm,内部断层扫描可以将X射线剂量减少18%至58%,并向探测器散射19%至59%。