Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound device.

机译:有晶状眼和假晶状体眼的前房深度测量:Pentacam与超声仪的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To compare anterior chamber depth (ACD) measurements with a new optical device with those taken with a standard ultrasound (US) device in emmetropic phakic and pseudophakic eyes. SETTING: Department of Ophthalmology, Medical Health and Science Center, University of Debrecen, Debrecen, Hungary. METHODS: Forty-two phakic and 42 pseudophakic patients with normal axial lengths (mean 22.91 mm +/- 1.21 [SD]) were enrolled in the study. The ACD was measured 3 times with Scheimpflug-based Pentacam (Oculus) and then 3 times with a standard A-scan US device (AL-2000, Tomey). The data were then analyzed. RESULTS: In the phakic group, the mean ACD was 2.87 +/- 0.4 mm with the Pentacam and 2.89 +/- 0.49 mm with ultrasound A-scan (US) (P = .84). In the pseudophakic group, the mean ACD was 3.41 +/- 0.28 mm and 3.97 +/- 0.45 mm, respectively (P < .001). The correlation between measurements was significant in both the phakic and pseudophakic groups (r = .547/P < .001 and r = .404/P = .01, respectively). CONCLUSIONS: In phakic eyes, ACD measured with the Pentacam and with US was the same. However, in pseudophakic eyes, the difference was significantly lower when the ACD was measured with the Pentacam. Therefore, in pseudophakic patients, further evaluation of ACD data with the Scheimpflug-based system is necessary.
机译:目的:比较正视性晶状体眼和假晶状体眼中使用新型光学设备的前房深度(ACD)测量值与使用标准超声(US)设备进行的前房深度测量值之间的比较。地点:匈牙利德布勒森德布勒森大学医学卫生与科学中心眼科。方法:本研究纳入了42例晶状体眼和42例假晶状体患者,这些患者的轴向长度正常(平均22.91 mm +/- 1.21 [SD])。使用基于Scheimpflug的Pentacam(Oculus)测量ACD 3次,然后使用标准A-scan US设备(AL-2000,Tomey)测量3次。然后分析数据。结果:在晶状体组中,Pentacam的平均ACD为2.87 +/- 0.4 mm,超声A扫描(US)的平均ACD为2.89 +/- 0.49 mm(P = 0.84)。在假晶状体组中,平均ACD分别为3.41 +/- 0.28 mm和3.97 +/- 0.45 mm(P <.001)。在晶状体和假晶状体组中,测量值之间的相关性均显着(r = .547 / P <.001和r = .404 / P = .01)。结论:在有晶状体眼中,用Pentacam和US测量的ACD相同。但是,在伪晶状体眼中,当使用Pentacam测量ACD时,差异显着降低。因此,对于假晶状体患者,必须使用基于Scheimpflug的系统进一步评估ACD数据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号