首页> 外文期刊>Journal of cataract and refractive surgery >Underestimation of corneal thickness by Orbscan after myopic correction. 15519055
【24h】

Underestimation of corneal thickness by Orbscan after myopic correction. 15519055

机译:近视矫正后Orbscan低估了角膜厚度。 15519055

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

In the reply1 to a letter about the article on corneal thickness measured by Orscan (Bausch & Lomb) ,2 Miyata could not satisfactorily explain why the corneal thickness was underestimated after myopic LASIK. In the article, Miyata et al. speculated that the reasons were changes in the refractive index, stromal haze, changes in the anterior contour of the cornea, and inappropriate reconstruction algorithms that might occur after LASIK. We propose another explanation of the underestimation of corneal thickness measured by Orbscan after myopic LASIK.The posterior surface of the cornea is observed through a lens composed of the overlying epithelium and stroma. The lens changes its shape and thickness after myopic LASIK. Then, the posterior surface of the cornea observed through the lens becomes relatively smaller than that observed before the surgery.
机译:在关于Orscan(Bausch&Lomb)测量的有关角膜厚度的文章的一封信的回复1中,Miyata不能令人满意地解释为什么近视LASIK后角膜厚度被低估了。在文章中,宫田等。推测其原因是折射率改变,基质雾度,角膜前轮廓的改变以及LASIK后可能发生的不合适的重建算法。我们提议对近视LASIK手术后Orbscan测量的角膜厚度低估的另一种解释。通过覆盖上皮和基质的晶状体观察角膜的后表面。近视LASIK术后晶状体会改变其形状和厚度。然后,通过晶状体观察到的角膜后表面变得比手术前观察到的相对小。