Comparison of methods to measure corneal power for intraocular lens power calculation using a rotating Scheimpflug camera

机译:使用旋转Scheimpflug相机测量角膜屈光度以计算人工晶状体屈光度的方法的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To assess the accuracy of corneal power measurements provided by a Scheimpflug camera (Pentacam HR) for intraocular lens (IOL) power calculation in unoperated eyes and compare the results with those of simulated keratometry (SimK) performed with a Placido-disk corneal topographer (Keratron). Setting: Private practice. Design: Evaluation of diagnostic test. Methods: Eight Scheimpflug camera corneal power measurements were analyzed: (1) average K, (2) true net power calculated using the Gaussian optics formula, (3) total corneal refractive power at 2.0 mm calculated by ray tracing on a ring and (4) as the average of the zone inside the ring, (5) total corneal refractive power at 3.0 mm on a ring and (6) as the average of the zone inside the ring, (7) the equivalent K reading at 3.0 mm and (8) at 4.5 mm. The IOL power was calculated using the Hoffer Q, Holladay 1, and SRK/T formulas. Results: No statistically significant differences were observed between any corneal power measurements, including simulated K, in 41 consecutive patients. The latter showed slightly lower mean absolute errors with all 3 formulas (range 0.26 to 0.27 diopter [D]). The Scheimpflug camera gave the lowest median absolute errors with all formulas; that is, the 3.0 mm equivalent K reading with the Hoffer Q formula (0.18 D) and Holladay 1 formula (0.17 D) and the 2.0 mm total corneal refractive power ring with the SRK/T formula (0.18 D). Conclusion: Corneal power measurements provided by the Scheimpflug camera and Placido disk corneal topographer displayed comparable accuracy in IOL power calculation. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.
机译:目的:评估Scheimpflug相机(Pentacam HR)提供的角膜屈光度测量的准确性,以在未手术的眼睛中计算人工晶状体(IOL)屈光度,并将结果与​​使用Placido盘状角膜地形图仪进行的模拟角膜测量法(SimK)进行比较(Keratron)。地点:私人执业。设计:诊断测试评估。方法:分析了八个Scheimpflug相机角膜屈光力测量值:(1)平均K,(2)使用高斯光学公式计算的真实净屈光力,(3)通过在圆环上进行光线追踪计算的2.0 mm的总角膜屈光力,以及(4 )作为环内区域的平均值,(5)环上3.0 mm处的总角膜屈光力,以及(6)作为环内区域的平均值,(7)3.0 mm处的等效K读数,以及( 8)在4.5毫米处。使用Hoffer Q,Holladay 1和SRK / T公式计算IOL功效。结果:连续41例患者的任何角膜屈光力测量值(包括模拟K值)之间均未观察到统计学上的显着差异。后者在所有3个公式中均显示出较低的平均绝对误差(0.26到0.27屈光度[D])。 Scheimpflug相机在所有公式中的绝对绝对值最低。也就是说,使用Hoffer Q公式(0.18 D)和Holladay 1公式(0.17 D)的3.0毫米等效K读数,以及使用SRK / T公式(0.18 D)的2.0毫米总角膜屈光力环。结论:Scheimpflug相机和Placido盘角膜地形图仪提供的角膜屈光度测量在IOL屈光度计算中显示出可比的准确性。财务披露:没有任何作者对所提及的任何材料或方法有财务或专有利益。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号