首页> 外文期刊>Clinical biomechanics >Accurate diagnosis of hip prosthesis loosening using a vibrational technique.
【24h】

Accurate diagnosis of hip prosthesis loosening using a vibrational technique.

机译:使用振动技术准确诊断髋关节假体松动。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To examine the potential role of vibration testing as a non-invasive method of diagnosing loosening of total hip replacements in the clinical setting. DESIGN: Single blind cohort study in two hospitals. BACKGROUND: Diagnosing loosening of total hip replacements is heavily dependent on investigative techniques that are unreliable. Previous studies into the use of vibration testing have produced conflicting results. METHODS: Comparison of vibration testing and radiographs in patients with a total hip replacement experiencing hip pain symptomatic of loosening, with patients showing evidence of a secure prosthesis. RESULTS: Vibration testing has a sensitivity of 80% and a specificity of 89%. The positive predictive value was 92% and the negative predictive value was 73%; it was unable to produce a definitive diagnosis in 8% of patients. When compared with radiographs from the same patients, vibration testing was shown to be 20% more sensitive and able to diagnose 13% more patients. CONCLUSIONS: Vibration testing can deliver more accurate information on the stability of total hip replacements than radiographs in the clinical setting, despite being in the early stages of development. Relevance. This study shows that 70 more patients may be provided with the correct diagnosis and 46 fewer patients may be undiagnosed each year, when using vibrometry as opposed to radiographs. In view of the relative disparity between the level of development between the two techniques and the encouraging results hitherto presented, it is felt that by improving vibration testing it may supersede radiographs in the detection of prosthesis loosening.
机译:目的:探讨振动测试作为一种非侵入性方法在临床中诊断全髋关节置换术松动的潜在作用。设计:在两家医院进行单盲队列研究。背景:诊断全髋关节置换术的松动在很大程度上取决于不可靠的调查技术。先前对振动测试的使用研究产生了矛盾的结果。方法:比较全髋关节置换患者出现髋关节疼痛症状松动的振动测试和X射线照片的比较,并显示假体安全的证据。结果:振动测试的灵敏度为80%,特异性为89%。阳性预测值为92%,阴性预测值为73%;它无法对8%的患者做出明确的诊断。与来自相同患者的X射线照片相比,振动测试的敏感性提高了20%,能够诊断出13%的患者。结论:尽管处于开发的早期阶段,但振动测试可以提供比临床上的X光片更全面的髋关节置换稳定性信息。关联。这项研究表明,与放射线照相相比,使用振动测定法时,每年可以为70多名患者提供正确的诊断,而每年有46名更少的未被诊断的患者。考虑到两种技术之间的发展水平和迄今为止提出的令人鼓舞的结果之间的相对差异,可以感觉到,通过改进振动测试,在假体松动的检测中可以取代射线照相。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号