Assessment of therapeutic options for mild obstructive sleep apnea using cardiopulmonary coupling measures

机译:使用心肺耦合措施评估轻度阻塞性睡眠呼吸暂停的治疗选择

获取原文
获取原文并翻译 | 示例
       

摘要

Objective: To examine the efficacy of various therapeutic modalities for mild obstructive sleep apnea using cardiopulmonary coupling variables of sleep quality. Methods: A 67-year-old Caucasian subject's sleep was recorded at home for 10 nights using a type 3 sleep recording device that measured ECG and body position, followed by generation of the cardiopulmonary sleep spectrogram. Three baseline nights, one night with a sleep jacket containing 3 tennis balls to restrict sleep in the supine position, 2 nights with oxygen only delivered via a nasal cannula at a fl ow rate of 2 L/minute, 2 nights with a mandible advancing appliance (MAA) only, and 2 nights using oxygen at 2 L/minute with the MAA were compared. Results: Baseline sleep quality estimated using the ratio of high-frequency and low-frequency coupling (1.03) was below the expected normal adult values ranging from 1.67-4.0. The sleep quality ratio was signifi cantly higher (2.08) using the MAA alone compared to baseline, sleep position restriction (1.61), oxygen therapy (0.81), and the combination of MAA with oxygen (1.66). Conclusion: Sleep quality measured objectively using cardiopulmonary coupling variables differentiated the effi cacy of therapeutic options for mild obstructive sleep apnea. Such an approach may have practical utility.
机译:目的:使用睡眠质量的心肺耦合变量,检查各种治疗方式对轻度阻塞性睡眠呼吸暂停的疗效。方法:使用3型睡眠记录设备在家里测量一个67岁的白种人受试者的睡眠10个晚上,该设备可以测量ECG和身体位置,然后生成心肺睡眠频谱图。三个基准夜晚,一个夜晚,一个装有3个网球的睡衣,以限制仰卧姿势睡眠;两个夜晚,仅以2 L / min的流量通过鼻导管输送氧气,两个夜晚,带下颌前移装置(MAA),并比较了以2 L / min的氧气与MAA使用2晚的情况。结果:使用高频和低频耦合之比(1.03)估算的基准睡眠质量低于预期的正常成年人值(1.67-4.0)。与基线,限制睡眠的位置(1.61),氧气疗法(0.81)和MAA与氧气的组合(1.66)相比,单独使用MAA的睡眠质量比率显着更高(2.08)。结论:使用心肺耦合变量客观测量的睡眠质量区分了轻度阻塞性睡眠呼吸暂停的治疗选择的有效性。这样的方法可能具有实用性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号