Implantable cardioverter defibrillator sensing failure due to endocardial R wave electrical alternans.

机译:植入式心脏复律除颤器感测由于心内膜R波电交替引起的故障。

获取原文
获取原文并翻译 | 示例

摘要

A 71-year-old patient underwent routine single-chamber implantable cardioverter defibrillator (ICD) generator replacement. During defibrillation threshold testing, ventricular tachycardia was induced but the ICD failed twice to properly detect the tachycardia due to endocardial R wave alternans and sensing of every other beat. The problem was resolved by inserting a separate sensing/pacing lead positioned further away from the existing ICD lead.