Exploding implantable cardioverter defibrillator.

机译:爆炸式植入式心脏复律除颤器。

获取原文
获取原文并翻译 | 示例
       

摘要

A 79-year-old man with a pectoral implantable cardioverter defibrillator (ICD) system underwent periodic defibrillation threshold testing 18 months after implant. Attempted delivery of a 15-J shock caused a light flash above the pocket and a loud "pop." High-voltage lead impedance was <20 ohms. Pocket exploration revealed insulation abrasion of the high-voltage portion of the single-coil right ventricular lead. The outer shield of the active can pulse was perforated and scorched due to arcing. Device analysis confirmed a shorted transistor in the high-voltage output circuit. Unsuspected physical contact between high-voltage electrodes of opposite polarity within the pocket can cause catastrophic ICD system failure.
机译:一名患有胸膜植入式心脏复律除颤器(ICD)系统的79岁男性在植入后18个月接受了定期除颤阈值测试。尝试传递15-J的电击会导致口袋上方的闪光灯闪烁并发出响亮的“砰砰”声。高压引线阻抗小于20欧姆。袋式探查发现单线圈右心室导线高压部分的绝缘磨损。主动罐脉冲的外部屏蔽层由于电弧而穿孔并烧焦。器件分析确认高压输出电路中的晶体管短路。口袋内极性相反的高压电极之间的意外物理接触会导致灾难性的ICD系统故障。