The surfaces of retrieved failed cementless total hip implants made of cobalt-chromium-molybdenum casting alloy and of wrought titanium 6-aluminum 4-vanadium alloy were studied with the use of scanning-electron microscopy (SEM), energy-dispersive X-ray analysis (EDX) and X-ray photoelectron spectroscopy (XPS). New implants of the same make served as controls. The XPS scans revealed a dense carbon layer on the entire analyzed specimen. The relative composition of the titanium alloy implants showed an overall agreement with the international standards for implants for surgery, and the overall surface composition did not change over the period of the implantation. However, an inhomogeneous distribution of the constituents could be demonstrated in the retrieved as well as in the new MEC-screw rings made of TiAl6V4 alloy, an implant that has been linked to a high early failure rate. In the CoCr-alloy components (Lord-screw rings) a high percentage of aluminum, mainly organized in aluminum inclusions, was found in the retrieved as well as in the new implants.
机译:利用扫描电子显微镜(SEM),能量色散X射线分析法研究了由钴-铬-钼铸造合金和变形的钛6-铝4-钒合金制成的失效的非骨水泥全髋关节植入物的表面(EDX)和X射线光电子能谱(XPS)。相同品牌的新植入物用作对照。 XPS扫描显示整个分析样品上有致密的碳层。钛合金植入物的相对组成与外科植入物的国际标准完全吻合,并且整个表面组成在植入期间没有变化。然而,在回收的以及由TiAl6V4合金制成的新型MEC螺钉环中,可以证明成分的不均匀分布,这种植入物与早期的高失败率有关。在回收的以及新植入物中发现了高含量的钴铬合金部件(主螺纹环)中的铝百分比很高,主要组织为铝夹杂物。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号