Wear remains an important cause of failure in knee replacement. Of the current methods of early performance assessment or prediction. simulators have been un-physiological, single X-ray film analyses remain limited by accuracy and retrieval and survival methods have a prohibitive time scale. An accurate method is needed to allow a timely assessment of polyethylene component wear in vivo. when a new design is introduced, in order to predict likely outcome.We present a new method for measuring wear in vivo that we believe will allow this prediction of long-term wear. X-ray film pairs were taken of implanted prosthetic metal components. When the X-ray system was calibrated, projections of the appropriate Computer Aided Design (CAD) model could be matched to the shapes on the scanned X-ray films to find component positions. Interpenetration of the metal femoral component into the polyethylene component could then be established and represents our estimate of "wear". This method was used to measure in vivo prosthesis wear to an accuracy of 0.11 mm. (C) 2004 Elsevier Ltd. All rights reserved.
机译:磨损仍然是导致膝关节置换失败的重要原因。当前的早期绩效评估或预测方法。模拟器是非生理性的,单次X射线胶片分析仍然受到准确性的限制,而检索和生存方法的时间尺度却令人望而却步。需要一种准确的方法以允许及时评估体内聚乙烯组分的磨损。在介绍新设计时,为了预测可能的结果。我们提出了一种新的体内磨损测量方法,我们相信这种方法可以预测长期磨损。 X射线胶片对取自植入的修复金属部件。校准X射线系统后,可以将适当的计算机辅助设计(CAD)模型的投影与扫描的X射线胶片上的形状进行匹配,以找到组件的位置。然后可以确定金属股骨组件与聚乙烯组件之间的相互渗透,这代表了我们对“磨损”的估计。该方法用于测量体内假体磨损,精确度为0.11 mm。 (C)2004 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号