首页> 外文期刊>Journal of Biomechanics >Comparison of methods for the calculation of energy storage and return in a dynamic elastic response prosthesis
【24h】

Comparison of methods for the calculation of energy storage and return in a dynamic elastic response prosthesis

机译:动态弹性响应假体中能量存储和返回的计算方法比较

获取原文
获取原文并翻译 | 示例
       

摘要

The standard method used to calculate the ankle joint power contains deficiencies when applied to dynamic elastic response prosthetic feet. The standard model, using rotational power and inverse dynamics, assumes a fixed joint center and cannot account for energy storage, dissipation, and return. This study compared the standard method with new analysis models. First, assumptions of inverse dynamics were avoided by directly measuring ankle forces and moments. Second, the ankle center of rotation was corrected by including translational power terms. Analysis with below-knee amputees revealed that the conventional method overestimates ankle forces and moments as well as prosthesis energy storage and return. Results for efficiency of energy return were varied. Large differences between models indicate the standard method may have serious inadequacies in the analysis of certain prosthetic feet. This research is the first application of the new models to prosthetic feet, and suggests the need for additional research in gait analysis with energy-storing prostheses. (C) 2000 Elsevier Science Ltd. All rights reserved. [References: 10]
机译:当应用于动态弹性响应假肢脚时,用于计算踝关节力量的标准方法存在缺陷。使用旋转功率和逆动力学的标准模型假定关节中心固定,并且无法考虑能量的存储,耗散和返回。这项研究将标准方法与新的分析模型进行了比较。首先,通过直接测量脚踝力和力矩避免了逆动力学的假设。其次,通过包括平移功率项来校正踝关节旋转中心。对膝盖以下截肢者的分析显示,传统方法过高估计了脚踝的力量和力矩以及假体的能量存储和返回。能量返回效率的结果各不相同。模型之间的较大差异表明标准方法在某些假脚的分析中可能存在严重不足。这项研究是新模型首次应用于假足的研究,并提出了在储能假肢的步态分析中需要进行其他研究的需求。 (C)2000 Elsevier ScienceLtd。保留所有权利。 [参考:10]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号