Using ground reaction forces recorded while propelling a manual wheelchair on an instrumented treadmill may represent a valuable alternative to using an instrumented pushrim to calculate temporal and kinetic parameters during propulsion. Sixteen manual wheelchair users propelled their wheelchair equipped with instrumented pushrims (i.e., SMARTWheel) on an instrumented dual-belt treadmill set a 1 m/s during a 1-minute period. Spatio-temporal (i.e., duration of the push and recovery phase) and kinetic measures (i.e. propulsive moments) were calculated for 20 consecutive strokes for each participant. Strong associations were confirmed between the treadmill and the instrumented pushrim for the mean duration of the push phase (r=0.98) and of the recovery phase (r=0.99). Good agreement between these two measurement instruments was also confirmed with mean differences of only 0.028 s for the push phase and 0.012 s for the recovery phase. Strong associations were confirmed between the instrumented wheelchair pushrim and treadmill for mean (r=0.97) and peak (r=0.96) propulsive moments. Good agreement between these two measurement instruments was also confirmed with mean differences of 0.60 Nm (mean moment) and 0.71 Nm (peak moment). The use of a dual-belt instrumented treadmill represents an alternative to characterizing temporal parameters and propulsive moments during manual wheelchair propulsion. (C) 2016 Elsevier Ltd. All rights reserved.
机译:使用在仪表式跑步机上推动手动轮椅时记录的地面反作用力,可能是在推进过程中使用仪表式轮辋计算时间和动力学参数的一种有价值的替代方法。 16名手动轮椅使用者在1分钟内以1 m / s的速度将装有器械轮缘(即SMARTWheel)的轮椅推进到器械双皮带跑步机上。为每个参与者计算连续20次笔划的时空(即推和恢复阶段的持续时间)和动力学指标(即推进力矩)。在推力阶段(r = 0.98)和恢复阶段(r = 0.99)的平均持续时间内,证实了跑步机和仪表式推圈之间的强关联。还证实了这两种测量仪器之间的良好一致性,推动阶段的平均差异仅为0.028 s,恢复阶段的平均差异仅为0.012 s。证实了仪器的轮椅前缘和跑步机之间的均值(r = 0.97)和峰值(r = 0.96)推进力矩之间有很强的联系。还证实了这两种测量仪器之间的良好一致性,其平均差为0.60 Nm(平均矩)和0.71 Nm(峰值矩)。使用双皮带仪表式跑步机是表征手动轮椅推进过程中的时间参数和推进力矩的一种替代方法。 (C)2016 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号