首页> 外文期刊>Journal of analytical chemistry >Iodide ion selective poly(aniline) solid contact electrode based on quinine-cu ionophores
【24h】

Iodide ion selective poly(aniline) solid contact electrode based on quinine-cu ionophores

机译:基于奎宁铜离子载体的碘离子选择性聚苯胺固体接触电极

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Iodide ion selective poly(aniline) solid contact electrode based on quinine-Cu ionophore as a sensing material has been successfully developed. The electrode exhibits good linear response of 52.0 mV/decade (at 20 ± 0.2°C, r ~2 = 0.9998) within the concentration range of 1 × 10~(-6.4)-1 × 10~(-1.0) M KI. The composition of this electrode was quinine-Cu 2.0: PVC 30.0: bis(2-ethylhexyl)sebacate 68.0 (mass). This plasticizer provides the best response characteristics. The electrode shows good selectivity for iodide ion in comparison with any other anions and is suitable for use with aqueous solutions of pH 3.3-9.4. The standard deviations of the measured EMF difference were ±1.4 and ±1.3 mV for iodide sample solutions of 1.0 × 10~(-2) M and 1.0 × 10 ~(-3) M, respectively. The stabilization time was less than 10 min and response time was less than 15 sec.
机译:以奎宁-铜离子载体为传感材料的碘化物离子选择性聚苯胺固体接触电极已成功开发。在1×10〜(-6.4)-1×10〜(-1.0)M KI的浓度范围内,该电极表现出良好的线性响应,为52.0 mV / decade(在20±0.2°C,r〜2 = 0.9998)。该电极的组成为奎宁-Cu 2.0:PVC 30.0:癸二酸双(2-乙基己基)酯68.0(质量)。该增塑剂具有最佳的响应特性。与其他阴离子相比,该电极对碘离子具有良好的选择性,并且适合与pH 3.3-9.4的水溶液一起使用。对于1.0×10〜(-2)M和1.0×10〜(-3)M的碘化物样品溶液,测得的EMF差的标准偏差分别为±1.4和±1.3 mV。稳定时间少于10分钟,响应时间少于15秒。